Abstract
The chromatographic characteristics were determined for a set of microfabricated separation channels structured with cylindrical and diamond‐shaped pillars with a characteristic size of 5 μm. Channels with different structures and porosities were etched in a silicon wafer using lithographic techniques. The permeability for flow of the channels was shown to increase strongly with the overall porosity. Diamond‐shaped pillars appeared to yield a slightly higher permeability than cylindrical pillars at the same channel porosity. Compared to packed columns, permeabilities were higher by a factor of up to 5. Band dispersion in the channels was measured with an unretained fluorescent probe compound using a fluorescence microscope. A relatively large variation in the observed plate heights between channels was found, which was mainly attributed to the inaccurate geometry of the structure close to the side walls. Reduced plate heights between 0.2 and 1.0 were obtained. The lowest plate heights were found for channels with low porosity. The chromatographic impedances were calculated and compared to the values for the traditional chromatographic systems. For one of the structured microchannels the impedance was found to be more than ten times lower than for a column packed with nonporous spherical particles. With the data collected, predictions are given on the possibilities in terms of efficiency and speed offered by structured microchannels for pressure‐driven separations, taking practical constraints into account.Keywords
This publication has 16 references indexed in Scilit:
- Experimental Study on Band Dispersion in Channels Structured with MicropillarsAnalytical Chemistry, 2006
- Tailoring the Morphology of Methacrylate Ester-Based Monoliths for Optimum Efficiency in Liquid ChromatographyAnalytical Chemistry, 2005
- General Rules for the Optimal External Porosity of LC SupportsAnalytical Chemistry, 2004
- Importance and Reduction of the Sidewall-Induced Band-Broadening Effect in Pressure-Driven Microfabricated ColumnsAnalytical Chemistry, 2004
- Influence of the Pillar Shape on the Band Broadening and the Separation Impedance of Perfectly Ordered 2-D Porous Chromatographic MediaAnalytical Chemistry, 2004
- Advantages of Perfectly Ordered 2-D Porous Pillar Arrays over Packed Bed Columns for LC Separations: A Theoretical AnalysisAnalytical Chemistry, 2003
- Peer Reviewed: Monolithic LC Columns2001
- Some reflections on speed and efficiency of modern chromatographic methodsJournal of Chromatography A, 1997
- Factors determining flow rate in chromatographic columnsChromatographia, 1981
- Standardization of test conditions for high performance liquid chromatography columnsChromatographia, 1977