Corneal Endothelial Cell Loss After Cataract Extraction by Using Ultrasound Phacoemulsification Versus a Fluid-based System
Authors:
- Louis HoffartLHLouis Hoffart
- Frederic ChavaneFrederic Chavane
- Bernard RidingsBRBernard Ridings
- John ConrathJCJohn Conrath
- 1 January 2008
- journal article
- research article
- Published by Wolters Kluwer Health in Cornea
- Vol. 27 (1) , 17-21
Abstract
To compare in vivo corneal endothelial cell loss (ECL) after fluid-based versus ultrasound phacoemulsification. In this prospective randomized study, 2 groups of 21 patients underwent phacoemulsification with ultrasound (group 1) or a fluid-based system (group 2). A "divide and conquer" method with capsular bag lens implantation was performed. Cataracts graded up to nuclear opalescence 4.9 in Lens Opacities Classification System III (LOCS III) only were included. Endothelial cell density (ECD), percent hexagonality, endothelial cell area (ECA), coefficient of variation in cell size, and central corneal pachymetry were measured preoperatively, 7 days, 1 month, and 3 months after surgery. Statistical analysis was performed by using a 2-tailed Student t test. Average nuclear opalescence was identical in both groups (P = 0.908). Mean ECD was 1867 +/- 451 and 2031 +/- 400 cells/mm at 3 months postoperatively. Mean ECL was 498 +/- 415 (20.6% +/- 17.1%) and 302 +/- 302 cells/mm (13.3% +/- 13.2%) at 7 days, 589 +/- 342 (24.3% +/- 14%) and 254 +/- 214 cells/mm (11.2% +/- 9.1%) at 1 month, and 555 +/- 340 (22.9% +/- 14%) and 247 +/- 208 cells/mm (10.9% +/- 9.1%) at 3 months in groups 1 and 2, respectively. ECL was significantly lower for group 1 than for group 2 (P < 0.001) after the first postoperative month. A significant increase in ECA was also observed at the first and third (P = 0.012) postoperative months. Our study showed significantly lower ECL after phacoemulsification for a fluid-based versus an ultrasound system. The fluid-based method was safer than conventional ultrasound in cataract surgery, with a nuclear opalescence graded up to 4.9 in the LOCS III classification.Keywords
This publication has 27 references indexed in Scilit:
- Corneal endothelial function after phacoemulsification using the fluid-based system compared to conventional ultrasound techniqueEye, 2006
- Aqualase for cataract extractionEye, 2005
- Comparison of Corneal Endothelial Cell Images From a Noncontact Specular Microscope and a Scanning Confocal MicroscopeCornea, 2005
- AquaLase: a new technology for cataract extractionCurrent Opinion in Ophthalmology, 2004
- The relationship between nuclear colour and opalescence on the LOCSIII scale and physical characteristics of cataract nucleiEye, 2002
- Zerkleinerung humaner Linsenkerne mittels WaterjetDer Ophthalmologe, 2002
- Comparative study of three semiautomated specular microscopesJournal of Cataract & Refractive Surgery, 1995
- The Lens Opacities Classification System IIIArchives of Ophthalmology (1950), 1993
- Divide and conquer nucleofractis phacoemulsification: Development and variationsJournal of Cataract & Refractive Surgery, 1991
- Endothelial Cell Density Determined by Specular Microscopy and Scanning Electron MicroscopyOphthalmology, 1979