Abstract
Objective. Prompt gamma (PG) detection enables beam range verification and margin reduction in proton therapy, but its effectiveness is limited by counting statistics. Spot boosting has been proposed to enhance the signal-to-noise ratio by increasing the number of protons in selected spots of a pencil beam scanning treatment plan. This work integrates PG monitoring into an multi-criteria optimization (MCO) framework to investigate trade-offs between proton counts in boosted spots and dosimetric metrics. Approach. Intensity modulated proton therapy plans for a prostate-tumor and a brain-tumor patient were optimized in OpenTPS. Dose, LET weighted dose (LETxdose) and PG distributions were simulated for each pencil beam using the fast Monte Carlo (MC) engine MCSquare. For each field, ten candidate spots were selected based on three metrics quantifying the quality of PG monitoring and an even spatial distribution across the clinical target volume. Initial pencil beam weights and corresponding dose and LETxdose distributions were imported into an MCO framework for re-optimization, simultaneously considering both dosimetric goals and boosting objectives. Main results. Boosting up to 5 ⋅ 10 8 protons/spot maintained the plan quality without violating constraints in the prostate case. In the brain case, spot boosting required moderate deviations from the initial target dose constraints. For both cases, only minor boosting gains were obtained by relaxing dose metrics for organs at risk. Significance. This study is the first to visualize spot boosting trade-offs via Pareto fronts to support an informed decision, integrating verification quality aspects into the treatment planning stage. The approach enables selection of plans that preserve dose and LETxdose objectives while enhancing PG signal in targeted spots for reliable in-vivo range verification. The framework is broadly applicable and can provide a strategy for embedding verification-driven objectives into treatment optimization beyond PG monitoring.Keywords
Funding Information
- DFG (441208898)
This publication has 30 references indexed in Scilit:
- Bi‐Level Multi‐Criteria Optimization to Include Linear Energy Transfer Into Proton Treatment PlanningJournal of Multi-Criteria Decision Analysis, 2025
- Clinical advances and challenges in dose constraints for organs at risk in proton therapy: A systematic review of the last decade's literaturePrecision Radiation Oncology, 2025
- Efficient Approximation Quality Computation for Sandwiching Algorithms for Convex Multicriteria OptimizationJournal of Optimization Theory and Applications, 2025
- Potential margin reduction in prostate cancer proton therapy with prompt gamma imaging for online treatment verificationPhysics and Imaging in Radiation Oncology, 2023
- Detectability of Anatomical Changes With Prompt-Gamma Imaging: First Systematic Evaluation of Clinical Application During Prostate-Cancer Proton TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2023
- Automatic detection and classification of treatment deviations in proton therapy from realistically simulated prompt gamma imaging dataMedical Physics, 2022
- Proton range uncertainty reduction benefits for skull base tumors in terms of normal tissue complication probability (NTCP) and healthy tissue dosesMedical Physics, 2021
- Classification of the source of treatment deviation in proton therapy using prompt‐gamma imaging informationMedical Physics, 2020
- Prompt Gamma Imaging for In Vivo Range Verification of Pencil Beam Scanning Proton TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2017
- Reoptimization of Intensity Modulated Proton Therapy Plans Based on Linear Energy TransferInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2016