Intraoperative fabrication of patient-specific moulded implants for skull reconstruction: single-centre experience of 28 cases
- 18 January 2014
- journal article
- Published by Springer Nature in Acta Neurochirurgica
- Vol. 156 (4) , 793-803
Abstract
Intraoperatively fabricated polymethylmethacrylate (PMMA) implants based on computer-designed moulds were used to improve cosmetic results after hard tissue replacement. To assess the implant’s cosmetic and functional results we performed both subjective and objective assessments. This retrospective analysis was performed using a cohort of 28 patients who received PMMA implants between February 2009 and March 2012. The cosmetic and functional results were assessed using a patient questionnaire. Furthermore an objective volumetric subtraction score (0–100) was applied and implant thickness, as well as gaps and tiers, were measured. Patients mainly judged their cosmetic result as “good”. Two of the 28 patients found their cosmetic result unfavourable. The functional result and stability was mainly judged to be good. Measurements of implant thickness showed a very high correlation with the thickness of the contralateral bone. Volumetric subtraction led to a median quality of 80 on a scale from 0 to 100. Median gaps around the margins of the implant were 1.5 mm parietally, 1.7 mm frontally and 3.5 mm fronto-orbitally, and median tiers were 1.2 mm, 0 mm and 0 mm respectively. The overall rate of surgical revisions was 10.7 % (three patients). Two patients suffered from wound healing disturbances (7.1 %). The overall complication rate was comparable to other reports in the literature. Implantation of intraoperatively fabricated patient-specific moulded implants is a cost-effective and safe technique leading to good clinical results with a low complication rate.Keywords
This publication has 20 references indexed in Scilit:
- Intraoperative template-molded bone flap reconstruction for patient-specific cranioplastyNeurosurgical Review, 2012
- Reconstruction of Skull DefectsJournal of Craniofacial Surgery, 2009
- Long-term results following titanium cranioplasty of large skull defectsNeurosurgical Focus, 2009
- Biomaterials in Craniofacial SurgeryJournal of Craniofacial Surgery, 2009
- Cost-Effective Patient-Specific Intraoperative Molded CranioplastyJournal of Craniofacial Surgery, 2008
- Cranioplasty for Large Skull Defects with PMMA (Polymethylmethacrylate) or Tutoplast® Processed Autogenic Bone GraftsCentral European Neurosurgery, 2007
- Statistical validation of image segmentation quality based on a spatial overlap index1Academic Radiology, 2004
- Bone autografting of the calvaria and craniofacial skeletonSurgical Neurology, 2003
- Correction of Large (>25 cm2) Cranial Defects with “Reinforced” Hydroxyapatite Cement: Technique and ComplicationsNeurosurgery, 2003
- Cranioplasty with Individual Carbon Fibre Reinforced Polymere (CFRP) Medical Grade Implants Based on CAD/CAM TechniqueActa Neurochirurgica, 2002