Magnesium Test: Universal and Ultra‐Sensitive Method for Measuring Reliability of Thin‐Film‐Encapsulated Bioelectronic Implants in Physiological Environment
Open Access
- 30 April 2024
- journal article
- research article
- Published by Wiley in Advanced Functional Materials
- Vol. 34 (32)
Abstract
Next‐generation bioelectronic implants require miniaturization, durability, and long‐term functionality. Thin film encapsulations, prepared with inorganic or hybrid organic/inorganic designs, are essential for ensuring protection, low water permeation, adaptability, and structural strength. It is equally important to precisely measure their barrier performance, especially for in vivo use, to ensure the manufacture of reliable bioelectronics. Current monitoring solutions are not adequate: they are bulky, lack sensitivity, and are incompatible with microfabricated devices. Here, a comprehensive method is introduced to quantify the permeability of thin‐film encapsulation for bioelectronic implants both in situ and in real time. This method relies on monitoring the electrical resistance of the Mg film, which experiences corrosion due to water permeation, leading to Mg hydrolysis. An analytical model is proposed that predicts and quantifies this permeation, and is adaptable for various types of encapsulations, including hybrid multilayers. An unprecedented ultra‐low detection limit of 3 × 10−8 g m−2 d−1 at room temperature is demonstrated and the monitoring approach is validated in vivo using polyimide and poly(dimethylsiloxane)‐coated bioelectronics.Keywords
Funding Information
- Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung (200021_182055)
- Innosuisse - Schweizerische Agentur für Innovationsförderung (45944.1 IP‐ENG)
This publication has 58 references indexed in Scilit:
- Optical Monitoring of Water Side Permeation in Thin Film EncapsulationAdvanced Materials, 2024
- Recent Advances in Encapsulation of Flexible Bioelectronic Implants: Materials, Technologies, and Characterization MethodsAdvanced Materials, 2022
- Extended Barrier Lifetime of Partially Cracked Organic/Inorganic Multilayers for Compliant Implantable ElectronicsSmall, 2021
- Conformal, Ultra‐thin Skin‐Contact‐Actuated Hybrid Piezo/Triboelectric Wearable Sensor Based on AlN and Parylene‐Encapsulated Elastomeric BlendAdvanced Functional Materials, 2021
- Soft, Implantable Bioelectronic Interfaces for Translational ResearchAdvanced Materials, 2020
- Thin, Transferred Layers of Silicon Dioxide and Silicon Nitride as Water and Ion Barriers for Implantable Flexible Electronic SystemsAdvanced Electronic Materials, 2017
- Improvements of Permeation Barrier Coatings Using Encapsulated Parylene Interlayers for Flexible Electronic ApplicationsPlasma Processes and Polymers, 2007
- Basal ganglia neural responses during behaviorally effective deep brain stimulation of the subthalamic nucleus in rats performing a treadmill locomotion testSynapse, 2006
- Aqueous degradation of polyimidesJournal of Applied Polymer Science, 1971
- A New, General Synthetic Method for the Preparation of Linear Poly‐p‐xylylenesJournal of Polymer Science Part A-1: Polymer Chemistry, 1966