Porous Coated Motion Cervical Disc Replacement: A Biomechanical, Histomorphometric, and Biologic Wear Analysis in a Caprine Model
Authors:
- Bryan W. CunninghamBCBryan W. Cunningham
- Paul C. McAfeePMPaul C. McAfee
- Seok Woo KimSKSeok Woo Kim
- John C. SefterJSJohn C. Sefter
- Andrew CappuccinoACAndrew Cappuccino
- Luiz PimentaLPLuiz Pimenta
- 1 July 2006
- journal article
- Published by Wolters Kluwer Health in Spine
- Vol. 31 (15) , 1666-1673
Abstract
The biomechanical, histopathologic, and histomorphometric characteristics of cervical disc replacement were assessed in a caprine animal model. To investigate the biomechanical, porous ingrowth, and histopathologic characteristics of the Porous Coated Motion (PCM) Cervical Disc replacement (Cervitech, Inc., Rockaway, NJ). As an alternative to anterior cervical interbody arthrodesis, an artificial cervical disc serves to replace the symptomatic degenerated disc, restore the functional biomechanical properties of the motion segment, and preserve neurologic function. There were 12 mature Nubian goats divided into 2 groups based on postoperative survival periods of 6 (n = 6) and 12 months (n = 6). Using an anterior surgical approach, a complete discectomy was performed at the C3-C4, followed by implantation of the PCM device. Functional outcomes of the disc prosthesis were based on computerized tomography (CT), multidirectional flexibility testing, undecalcified histology, histomorphometric, and immunocytochemical analyses. There was no evidence of prosthesis loosening, or neurologic or vascular complications. CT showed the ability to image and assess the cervical spinal canal for the presence of compressive pathology in the area of the CoCrMo prosthesis. Multidirectional flexibility testing under axial rotation and lateral bending indicated no differences in the full range of intervertebral motion between the disc prosthesis and nonoperative controls (P > 0.05). Based on immunohistochemical and histologic analysis, there was no evidence of particulate debris, cytokines, or cellular apoptosis within the local or systemic tissues. Moreover, review of the spinal cord at the operative levels indicated no evidence of cord lesions, inflammatory reaction, wear particles, or significant pathologic changes in any treatment. Histomorphometric analysis at the metal-bone interface indicated the mean trabecular ingrowth of 40.5% +/- 24.4% and 58.65% +/- 28.04% for the 6 and 12-month treatments, respectively. To our knowledge, this serves as the first in vivo time-course study investigating the use of the PCM device for cervical arthroplasty. All 12 animals undergoing cervical disc replacement had no evidence of implant loosening, subluxation, or inflammatory reactions. PCM cervical arthroplasty permits unobstructed visualization of the spinal canal based on CT imaging. Segmental intervertebral motion was preserved under axial rotation and lateral bending loading conditions, while at the same time permitting porous osseointegration at the prosthesis-bone interface. Based on histopathologic review of all local and systemic tissues, there was no evidence of particulate wear debris, cytokines, cellular apoptosis, or significant pathologic changes in any treatment.Keywords
This publication has 14 references indexed in Scilit:
- Clinical experience with the new artificial cervical PCM (Cervitech) discThe Spine Journal, 2004
- Choosing a cervical disc replacementThe Spine Journal, 2004
- Cervical Disc Replacement—Porous Coated Motion ProsthesisSpine, 2003
- The effect of spinal instrumentation particulate wear debrisThe Spine Journal, 2003
- The Biologic Response to Particles From a Lumbar Disc ProsthesisSpine, 2002
- Measurement of spinal canal diameters in young subjects with lumbosacral transitional vertebraEuropean Spine Journal, 2001
- Anterior instability of the glenohumeral joint with humeral avulsion of the glenohumeral ligamentThe Journal of Bone and Joint Surgery. British volume, 1999
- Bone ingrowth and wear debris in well-fixed cementless porous-coated tibial components removed from patientsThe Journal of Arthroplasty, 1995
- Wear Debris in Total Joint ReplacementsJournal of the American Academy of Orthopaedic Surgeons, 1994
- A quantitative study of bone and soft tissues in cementless porous-coated acetabular components retrieved at autopsyThe Journal of Arthroplasty, 1993