Study protocol: stability of morphine injected without preservative, delivered with a disposable infusion device

Authors:

Abstract
Background and objective : Morphine hydrochloride, a major analgesic drug, is being increasingly administered using portable disposable infusion devices. The objective of this study was to investigate the stability of morphine in such a system at two concentrations (2·50 and 5·00 mg/ml) over a 30-day period.Method: High-performance liquid chromatography of stored morphine solutions.Results: The best stability was observed with disposable infusion devices filled with a morphine solution containing sodium metabisulphite as a preservative. No breakdown products were detected after 1 month of storage at room temperature, in light or darkness. On the other hand, 2·50 and 5·00 mg/ml morphine solutions without sodium metabisulphite, stored in the infusion device led to the formation of 0·205% and 0·235% of pseudomorphine, respectively, after 6 days of storage in the light, and 1·50% and 0·94% after 30 days storage.Conclusion: Morphine hydrochloride solutions stored in disposable infusion devices degraded very slowly, particularly when preserved with sodium metabisulphite. The solutions are stable over 5 days, the maximum period of storage normally required when using disposable infusers.