Accuracy of phase contrast, black‐blood, and bright‐blood pulse sequences for measuring compliance and distensibility coefficients in a human‐tissue mimicking phantom
Open Access
- 20 December 2009
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 31 (1) , 160-167
Abstract
Purpose: To assess the accuracy of MR‐derived luminal diameter variations and its implications for compliance (CC) and distensibility coefficients (DC) by comparison with high‐resolution digital photography (HRDP) in a tissue‐mimicking phantom with pulsatile flow.Materials and Methods: Diameters, CC, and DC extracted using cine phase‐contrast (CPC), cine bright‐blood (CBrB), and a cine black‐blood (CBB) sequence were compared. The diameter in the left–right direction was compared against HRDP, as the gold‐standard. The experiments were performed using 2562 and 5122 matrix sizes. Bland‐Altman analysis was performed to compare each sequence with the gold‐standard in terms of diameter changes over the simulated cardiac cycle.Results: The bias and 95% limits of agreement (LOA) for CBB and CBrB were comparable. The bias for CPC was larger, however, the LOA were comparable. Increasing spatial resolution improved agreement with HRDP for all sequences. CBrB‐derived CC and DC were within 3% of the high resolution CBB values while CPC CC and DC were underestimated but still within 11%.Conclusion: CPC images were found to underestimate the luminal area over the cardiac cycle. CBrB‐derived diameters were more accurate in diastole while CBB‐derived diameters gave the best results in systole. CC and DC varied depending on the pulse sequence. J. Magn. Reson. Imaging 2010;31:160–167.Keywords
This publication has 33 references indexed in Scilit:
- Relation of Aortic Wall Thickness and Distensibility to Cardiovascular Risk Factors (from the Multi-Ethnic Study of Atherosclerosis [MESA])The American Journal of Cardiology, 2008
- Development of a Composite Material Phantom Mimicking the Magnetic Resonance Parameters of the Neonatal Brain at 3.0 TeslaInvestigative Radiology, 2007
- Aging and Arterial Structure-Function Relations2006
- Blood Pressure, Large Arteries and Atherosclerosis2006
- Diabetes and Arterial Stiffening2006
- Black blood gradient echo cine magnetic resonance imaging of the mouse heartMagnetic Resonance in Medicine, 2005
- Method to correct for the effects of limited spatial resolution in phase‐contrast flow MRI measurementsMagnetic Resonance in Medicine, 2002
- Short and long-term effects of smoking on arterial wall properties in habitual smokersJACC, 1993
- Analysis of systematic and random error in MR volumetric flow measurementsMagnetic Resonance in Medicine, 1993
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986