Planar Biaxial Behavior of Fibrin-Based Tissue-Engineered Heart Valve Leaflets
- 1 October 2009
- journal article
- research article
- Published by SAGE Publications in Tissue Engineering Part A
- Vol. 15 (10) , 2763-2772
Abstract
To design more effective tissue-engineered heart valve replacements, the replacement tissue may need to mimic the biaxial stress–strain behavior of native heart valve tissue. This study characterized the planar biaxial properties of tissue-engineered valve leaflets and native aortic valve leaflets. Fibrin-based valve equivalent (VE) and porcine aortic valve (PAV) leaflets were subjected to incremental biaxial stress relaxation testing, during which fiber alignments were measured, over a range of strain ratios. Results showed that VE leaflets exhibited a modulus and fiber reorientation behavior that correlated with strain ratio. In contrast, PAV leaflets maintained their relaxed modulus and fiber alignment when exposed to nonequibiaxial strain, but exhibited changes in stress relaxation. In uniaxial and equi-biaxial tension, there were few observed differences in relaxation behavior between VE and PAV leaflets, despite differences in the modulus and fiber reorientation. Likewise, in both tissues there was similar relaxation response in the circumferential and radial directions in biaxial tension, despite different moduli in these two directions. This study presents some fundamental differences in the mechanical response to biaxial tension of fibrin-based tissue-engineered constructs and native valve tissue. It also highlights the importance of using a range of strain ratios when generating mechanical property data for valvular and engineered tissues. The data presented on the stress–strain, relaxation, and fiber reorientation of VE tissue will be useful in future efforts to mathematically model and improve fibrin-based tissue-engineered constructs.Keywords
This publication has 43 references indexed in Scilit:
- Time-dependent biaxial mechanical behavior of the aortic heart valve leafletJournal of Biomechanics, 2007
- Planar Biaxial Creep and Stress Relaxation of the Mitral Valve Anterior LeafletAnnals of Biomedical Engineering, 2006
- Autologous Human Tissue-Engineered Heart ValvesCirculation, 2006
- The Relation Between Collagen Fibril Kinematics and Mechanical Properties in the Mitral Valve Anterior LeafletJournal of Biomechanical Engineering, 2006
- Stability and function of glycosaminoglycans in porcine bioprosthetic heart valvesBiomaterials, 2006
- Biaixal Stress–Stretch Behavior of the Mitral Valve Anterior Leaflet at Physiologic Strain RatesAnnals of Biomedical Engineering, 2006
- Effect of sample geometry on the apparent biaxial mechanical behaviour of planar connective tissuesBiomaterials, 2005
- Loss of chondroitin 6‐sulfate and hyaluronan from failed porcine bioprosthetic valvesJournal of Biomedical Materials Research Part A, 2003
- Passive Biaxial Mechanical Properties of the Rat Bladder Wall After Spinal Cord InjuryJournal of Urology, 2002
- Quantification of strains in biaxially tested soft tissuesJournal of Biomechanics, 1987