Simultaneous Immunoassay Analysis of Plasma IL-6 and TNF-α on a Microchip
Open Access
- 9 January 2013
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 8 (1) , e53620
Abstract
Sandwich enzyme-linked immunosorbant assay (ELISA) using a 96-well plate is frequently employed for clinical diagnosis, but is time-and sample-consuming. To overcome these drawbacks, we performed a sandwich ELISA on a microchip. The microchip was made of cyclic olefin copolymer with 4 straight microchannels. For the construction of the sandwich ELISA for interleukin-6 (IL-6) or tumor necrosis factor-α (TNF-α), we used a piezoelectric inkjet printing system for the deposition and fixation of the 1st anti-IL-6 antibody or 1st anti-TNF-α antibody on the surface of the each microchannel. After the infusion of 2 µl of sample to the microchannel and a 20 min incubation, 2 µl of biotinylated 2nd antibody for either antigen was infused and a 10 min incubation. Then 2 µl of avidin-horseradish peroxidase was infused; and after a 5 min incubation, the substrate for peroxidase was infused, and the luminescence intensity was measured. Calibration curves were obtained between the concentration and luminescence intensity over the range of 0 to 32 pg/ml (IL-6: R2 = 0.9994, TNF-α: R2 = 0.9977), and the detection limit for each protein was 0.28 pg/ml and 0.46 pg/ml, respectively. Blood IL-6 and TNF-α concentrations of 5 subjects estimated from the microchip data were compared with results obtained by the conventional method, good correlations were observed between the methods according to linear regression analysis (IL-6: R2 = 0.9954, TNF-α: R2 = 0.9928). The reproducibility of the presented assay for the determination of the blood IL-6 and TNF-α concentration was comparable to that obtained with the 96-well plate. Simultaneous detection of blood IL-6 and TNF-α was possible by the deposition and fixation of each 1st antibody on the surface of a separate microchannel. This assay enabled us to determine simultaneously blood IL-6 and TNF-α with accuracy, satisfactory sensitivity, time saving ability, and low consumption of sample and reagents, and will be applicable to clinic diagnosis.Keywords
This publication has 27 references indexed in Scilit:
- Inflammatory markers in population studies of agingAgeing Research Reviews, 2011
- Quantitative Analysis of Serum Procollagen Type I C-Terminal Propeptide by Immunoassay on MicrochipPLOS ONE, 2011
- Marcadores inflamatórios da doença cardiovascular em idososArquivos Brasileiros de Cardiologia, 2009
- Piezoelectric inkjet printing of a cross‐hatch immunoassay on a disposable nylon membraneBiotechnology Journal, 2009
- Microchip-based cell analysis and clinical diagnosis systemLab on a Chip, 2008
- Guidelines for point‐of‐care testing: haematologyBritish Journal of Haematology, 2008
- Accurate quantitation of salivary and pancreatic amylase activities in human plasma by microchip electrophoretic separation of the substrates and hydrolysates coupled with immunoinhibitionElectrophoresis, 2008
- Demonstration of multi-analyte patterning using piezoelectric inkjet printing of multiple layersAnalytica Chimica Acta, 2008
- Microchip-based enzyme-linked immunosorbent assay (microELISA) system with thermal lens detectionLab on a Chip, 2004
- Determination of Carcinoembryonic Antigen in Human Sera by Integrated Bead-Bed Immunoasay in a Microchip for Cancer DiagnosisAnalytical Chemistry, 2001