The feasibility of determining the position of an endotracheal tube in neonates by using bedside ultrasonography compared with chest radiographs
Open Access
- 22 January 2015
- journal article
- Published by South African Medical Association NPC in South African Journal of Child Health
- Vol. 9 (1) , 3
Abstract
Background. Neonates in our neonatal intensive care unit (NICU) receive a large amount of radiation with X-rays (XRs) being done daily, even more often with reintubation, repositioning of endotracheal tubes (ETTs) and confirmation thereof, which has been our NICU policy for many years. Objective. To investigate the feasibility of determining the position of ETTs in neonates by using bedside ultrasonography (BUS), and to compare the results with those obtained from chest XR (CXR) findings. Methods. A prospective, cross-sectional study was done on intubated neonates in the NICU at Universitas Academic Hospital, Bloemfontein, to determine the position of ETTs by using BUS. Results. Thirty intubated patients included in this study had a median age of 13.5 days and a median weight of 1.6 kg. Ninety-three per cent of ETT placements were considered optimal when visualised by BUS, while 73.3% were considered to be placed optimally when CXR was viewed. When CXR and BUS findings were compared regarding optimal placing, the agreement was poor (κ=0.10; 95% confidence interval –0.2 - 0.4). In four patients, the distance from the aortic arch to the tip of the ETT was outside the expected range of 1.5 - 2.2 cm: in two patients it was 2.25 cm (6.7%). BUS measurements were done mainly in extended head (53.3%) or neutral (36.7%) position. Conclusion. Although poor agreement between CXR and BUS findings was obtained, possibly because of handling of patients with secondary shifting of ETTs, BUS was found not to be comparable with CXR, but an alternative feasible method to determine the optimal position of ETTs in the trachea in neonates when using other reference points, with the added advantage of no radiation exposure.Keywords
This publication has 10 references indexed in Scilit:
- Techniques to ascertain correct endotracheal tube placement in neonatesCochrane Database of Systematic Reviews, 2014
- A feasibility study on bedside upper airway ultrasonography compared to waveform capnography for verifying endotracheal tube location after intubationThe Ultrasound Journal, 2013
- Secondary confirmation of endotracheal tube position by diaphragm motion in right subcostal ultrasound viewInternational Journal of Critical Illness and Injury Science, 2013
- Ultrasound Confirmation of Endotracheal Tube Position in NeonatesNeonatology, 2012
- A Prospective Comparison of Diaphragmatic Ultrasound and Chest Radiography to Determine Endotracheal Tube Position in a Pediatric Emergency DepartmentPediatrics, 2009
- Use of Bedside Ultrasonography for Endotracheal Tube Placement in Pediatric Patients: A Feasibility StudyPediatrics, 2007
- Optimal Endotracheal Tube Tip Position in Extremely Premature InfantsAmerican Journal of Perinatology, 2007
- Secondary confirmation of endotracheal tube position by ultrasound imageCritical Care Medicine, 2004
- Effect of neck position on endotracheal tube location in low birth weight infantsPediatric Pulmonology, 1999
- First thoracic vertebral body as reference for endotracheal tube placement.Archives of Disease in Childhood - Fetal and Neonatal Edition, 1994