Abstract
Purpose: O3® Regional Oximetry (Masimo Corporation, California, USA) is validated for cerebral oximetry. We aimed to assess agreement of somatic and renal near-infrared spectroscopy with reference blood samples. Methods: O3 sensors were placed bilaterally on the quadriceps and flank of 26 healthy volunteers. A stepped, controlled hypoxia sequence was performed by adding a mixture of nitrogen and room air to the breathing circuit. O3-derived oxygen saturation values were obtained at baseline and at six decremental saturation levels (5% steps). Blood samples (radial artery, iliac vein (somatic reference) and renal vein) were obtained at each step. Reference values were calculated as: 0.7 × venous saturation + 0.3 × arterial saturation. The agreement between O3-derived values with blood reference values was assessed by calculating root-mean-square error accuracy and Bland-Altman plots. Results: The root-mean-square error accuracy was 6.0% between quadriceps oxygen saturation and somatic reference values. The mean bias was 0.8%, with limits of agreement from -7.7 to 9.3%. These were 5.1% and 0.6% (-8.3 to 9.5%) for flank oxygen saturation and somatic reference values, respectively, and 7.7% and -4.9% (-15.0 to 5.2%) for flank oxygen saturation and renal reference values. The kidney depth was 3.1 ± 0.9 cm below the skin. Conclusion: O3 regional oximetry can be used on the quadriceps and flank to monitor somatic saturation, yet has a saturation-level dependent bias. O3-derived values obtained at the flank underestimated renal reference values. Additionally, it is unlikely that the flank sensors did directly measure renal tissue. Trial registration: Clinicaltrials.gov (NCT04584788): registered October 6th, 2020.Keywords
Funding Information
- Masimo Corporation
This publication has 32 references indexed in Scilit:
- Intraoperative tissue oxygenation and postoperative outcomes after major non-cardiac surgery: an observational studyBritish Journal of Anaesthesia, 2013
- Near infrared spectroscopy (NIRS) of the thenar eminence in anesthesia and intensive careAnnals of Intensive Care, 2012
- Near-infrared spectroscopy: A methodology-focused reviewSeminars in Fetal and Neonatal Medicine, 2011
- Confidence in Altman–Bland plots: A critical review of the method of differencesClinical and Experimental Pharmacology and Physiology, 2010
- Acute Kidney Injury Is Associated With Increased Long-Term Mortality After Cardiothoracic SurgeryCirculation, 2009
- The Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) Statement: Guidelines for Reporting Observational StudiesAnnals of Internal Medicine, 2007
- DYNAMIC NEAR-INFRARED SPECTROSCOPY MEASUREMENTS IN PATIENTS WITH SEVERE SEPSISShock, 2007
- NEAR-INFRARED SPECTROSCOPYCritical Care Clinics, 1996
- Role of Oxygen Debt in the Development of Organ Failure Sepsis, and Death in High-Risk Surgical PatientsChest, 1992
- Tissue oxygen debt as a determinant of lethal and nonlethal postoperative organ failureCritical Care Medicine, 1988