3D myocardial T1 mapping at 3T using variable flip angle method: Pilot study
Open Access
- 25 February 2013
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 71 (2) , 823-829
Abstract
Purpose: Myocardial T1 mapping is an emerging technique that could improve cardiovascular magnetic resonance diagnostic accuracy. In this study, a variable flip angle approach with B1 correction is proposed at 3T on the myocardium, employing standard 3D spoiled fast gradient echo and echo planar imaging sequences.Methods: The method was tested on phantoms to determine the set of standard 3D spoiled fast gradient echo angles adapted to myocardial T1 measurements and was compared to the inversion‐recovery spin‐echo reference T1 method. Seven volunteers underwent magnetic imaging resonance to acquire myocardial T1 maps and T1 values of the human heart.Results: This original method demonstrated good reproducibility in phantoms and a significant correlation between variable flip angle T1 values and reference inversion‐recovery spin‐echo T1 values. It yielded myocardial T1 values consistent with expected T1 and an increasing homogenization of myocardial segments owing to B1 correction. The mean myocardial T1 value was 1341 ± 42 ms.Conclusion: Myocardial 3D T1 mapping using the variable flip angle approach can potentially be useful for evaluating fibrosis on the entire myocardium using a standard clinical sequence. Magn Reson Med 71:823–829, 2014.Keywords
Funding Information
- the Région Lorraine
This publication has 29 references indexed in Scilit:
- T1 mapping of the myocardium: Intra-individual assessment of the effect of field strength, cardiac cycle and variation by myocardial regionJournal of Cardiovascular Magnetic Resonance, 2012
- Myocardial T1 mapping with MRI: Comparison of look‐locker and MOLLI sequencesJournal of Magnetic Resonance Imaging, 2011
- Rapid cardiac T1 mapping within two heartbeatsJournal of Cardiovascular Magnetic Resonance, 2011
- Myocardial T1 and extracellular volume fraction mapping at 3 teslaJournal of Cardiovascular Magnetic Resonance, 2011
- Shortened Modified Look-Locker Inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 1.5 and 3 T within a 9 heartbeat breathholdJournal of Cardiovascular Magnetic Resonance, 2010
- B1 mapping by Bloch‐Siegert shiftMagnetic Resonance in Medicine, 2010
- Measurement and characterization of RF nonuniformity over the heart at 3T using body coil transmissionJournal of Magnetic Resonance Imaging, 2008
- Cardiovascular Magnetic Resonance and prognosis in cardiac amyloidosisJournal of Cardiovascular Magnetic Resonance, 2008
- Myocardial T1 Mapping for Detection of Left Ventricular Myocardial Fibrosis in Chronic Aortic Regurgitation: Pilot StudyAmerican Journal of Roentgenology, 2006
- Modified Look‐Locker inversion recovery (MOLLI) for high‐resolution T1 mapping of the heartMagnetic Resonance in Medicine, 2004