Guard-Gap Correction for Guarded-Electrode Measurements and Exact Equations for the Two-Fluid Method of Measuring Permittivity and Loss
Author:
- 1 May 1976
- journal article
- Published by ASTM International in Journal of Testing and Evaluation
- Vol. 4 (3) , 188-195
Abstract
Empirical equations are developed for evaluating the guard-gap correction for guarded electrodes, taking into account the ratio of gap width to electrode spacing, the ratio of the electrode thickness to the gap width, and the permittivities of the material in the gap and between the electrodes. Exact equations are then developed for calculating specimen permittivity, thickness, and dissipation factor from two-fluid measurements, for the first time taking into account the change in effective area when the specimen is inserted between the electrodes, the dissipation factor of the specimen, and that of the liquid. Series resistance and inductance and change in spacing of the electrodes between the first and second set of readings are also taken into account. The accuracy of the equations is estimated to be such that they will not significantly affect the accuracy of two-fluid measurements. Sample calculations are included.Keywords
This publication has 5 references indexed in Scilit:
- Precise determination of the area of guarded electrodes for accurate dielectric measurements on solid-disk specimens1963
- The effective area of a guarded electrode1963
- Precise measurement of dielectric constant by the two-fluid technique1962
- Measurement of permittivity and dissipation factor without attached electrodes1960
- Measurement of the dielectric constants of thin films1950