In vitro validation and comparison of different software packages or algorithms for coronary bifurcation analysis using calibrated phantoms: Implications for clinical practice and research of bifurcation stenting
- 24 July 2014
- journal article
- research article
- Published by Wiley in Catheterization and Cardiovascular Interventions
- Vol. 85 (4) , 554-563
Abstract
Background: The accuracy and precision of quantitative coronary angiography (QCA) software dedicated for bifurcation lesions compared with conventional single‐vessel analysis remains unknown. Furthermore, comparison of different bifurcation analysis algorithms has not been performed.Methods: Six plexiglas phantoms with 18 bifurcations were manufactured with a tolerance < 10 µm. The bifurcation angiograms were analyzed using Cardiovascular Angiography Analysis System (CAAS; Version 5.10, Pie Medical Imaging, Maastricht, The Netherlands) and QAngio XA (Version 7.3, Medis Medical Imaging System BV, Leiden, The Netherlands) software packages.Results: Conventional single‐vessel analysis underestimated the reference vessel diameter and percent diameter stenosis in the proximal main vessel while it overestimated these parameters in the distal main vessel and side branch. CAAS software showed better overall accuracy and precision than QAngio XA (with automatic Y‐ or T‐shape bifurcation algorithm selection) for various phantom diameters including minimum lumen diameter (0.012 ± 0.103 mm vs. 0.041 ± 0.322 mm, P = 0.003), reference vessel diameter (−0.050 ± 0.043 mm vs. 0.116 ± 0.610 mm, P = 0.026), and % diameter stenosis (−0.94 ± 4.07 % vs. 1.74 ± 7.49 %, P = 0.041). QAngio XA demonstrated higher minimal lumen diameter, reference vessel diameter, and % diameter stenosis when compared to the actual phantom diameters; however, the accuracy of these parameters improved to a similar level as CAAS when the sole T‐shape algorithm in the QAnxio XA was used.Conclusion: The use of the single‐vessel QCA method is inaccurate in bifurcation lesions. Both CAAS and QAngio XA (when the T shape is systematically used) bifurcation software packages are suitable for quantitative assessment of bifurcations. © 2014 Wiley Periodicals, Inc.Keywords
This publication has 25 references indexed in Scilit:
- Advances in two-dimensional quantitative coronary angiographic assessment of bifurcation lesions: improved small lumen diameter detection and automatic reference vessel diameter derivationEuroIntervention, 2012
- Validity and variability in visual assessment of stenosis severity in phantom bifurcation lesions: A survey in experts during the fifth meeting of the european bifurcation clubCatheterization and Cardiovascular Interventions, 2011
- Novel bifurcation phantoms for validation of quantitative coronary angiography algorithmsCatheterization and Cardiovascular Interventions, 2011
- Dedicated bifurcation analysis: basic principlesThe International Journal of Cardiovascular Imaging, 2011
- Quantitative coronary analysis in the Nordic Bifurcation studiesThe International Journal of Cardiovascular Imaging, 2010
- Randomized Trial of Simple Versus Complex Drug-Eluting Stenting for Bifurcation LesionsCirculation, 2010
- Randomized comparison of everolimus- and paclitaxel-eluting stents: pooled analysis of the 2-year clinical follow-up from the SPIRIT II and III trialsEuropean Heart Journal, 2010
- Interventional Therapy of Bifurcation LesionsCirculation: Cardiovascular Interventions, 2009
- Immediate and long-term results of “T” stenting for bifurcation coronary lesionsThe American Journal of Cardiology, 2000
- In vivo validation of an experimental adaptive quantitative coronary angiography algorithm to circumvent overestimation of small luminal diametersCatheterization and Cardiovascular Diagnosis, 1995