Dose differences in intensity‐modulated radiotherapy plans calculated with pencil beam and Monte Carlo for lung SBRT
Open Access
- 8 November 2015
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 16 (6) , 91-99
Abstract
For patients with medically inoperable early-stage non-small cell lung cancer (NSCLC) treated with stereotactic body radiation therapy, early treatment plans were based on a simpler dose calculation algorithm, the pencil beam (PB) calculation. Because these patients had the longest treatment follow-up, identifying dose differences between the PB calculated dose and Monte Carlo calculated dose is clinically important for understanding of treatment outcomes. Previous studies found significant dose differences between the PB dose calculation and more accurate dose calculation algorithms, such as convolution-based or Monte Carlo (MC), mostly for three-dimensional conformal radiotherapy (3D CRT) plans. The aim of this study is to investigate whether these observed dose differences also exist for intensity-modulated radiotherapy (IMRT) plans for both centrally and peripherally located tumors. Seventy patients (35 central and 35 peripheral) were retrospectively selected for this study. The clinical IMRT plans that were initially calculated with the PB algorithm were recalculated with the MC algorithm. Among these paired plans, dosimetric parameters were compared for the targets and critical organs. When compared to MC calculation, PB calculation overestimated doses to the planning target volumes (PTVs) of central and peripheral tumors with different magnitudes. The doses to 95% of the central and peripheral PTVs were overestimated by and , respectively. This dose overestimation did not affect doses to the critical organs, such as the spinal cord and lung. In conclusion, for NSCLC treated with IMRT, dose differences between the PB and MC calculations were different from that of 3D CRT. No significant dose differences in critical organs were observed between the two calculations. PACS number: 87.53.Ly
Keywords
This publication has 15 references indexed in Scilit:
- Dosimetric errors during treatment of centrally located lung tumors with stereotactic body radiation therapy: Monte Carlo evaluation of tissue inhomogeneity correctionsMedical Dosimetry, 2013
- Dose calculation differences between Monte Carlo and pencil beam depend on the tumor locations and volumes for lung stereotactic body radiation therapyJournal of Applied Clinical Medical Physics, 2013
- Verification measurements and clinical evaluation of the iPlan RT Monte Carlo dose algorithm for 6 MV photon energyPhysics in Medicine & Biology, 2010
- Dosimetric verification and clinical evaluation of a new commercially available Monte Carlo-based dose algorithm for application in stereotactic body radiation therapy (SBRT) treatment planningPhysics in Medicine & Biology, 2010
- Intensity-Modulated Radiotherapy–Based Stereotactic Body Radiotherapy for Medically Inoperable Early-Stage Lung Cancer: Excellent Local ControlInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2010
- Experimental verification of a commercial Monte Carlo-based dose calculation module for high-energy photon beamsPhysics in Medicine & Biology, 2009
- A Comparison of Two Stereotactic Body Radiation Fractionation Schedules for Medically Inoperable Stage I Non-small Cell Lung Cancer: The Cleveland Clinic ExperienceJournal of Thoracic Oncology, 2009
- Dosimetric Evaluation of Heterogeneity Corrections for RTOG 0236: Stereotactic Body Radiotherapy of Inoperable Stage I-II Non–Small-Cell Lung CancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Stereotactic body radiation therapy of early-stage non–small-cell lung carcinoma: Phase I studyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2005
- Fast Monte Carlo dose calculation for photon beams based on the VMC electron algorithmMedical Physics, 1999