The impact of visual axis position on the optical quality after implantation of multifocal intraocular lenses with different asphericity values
Authors:
- 20 January 2021
- journal article
- research article
- Published by Springer Nature in Graefe's Archive for Clinical and Experimental Ophthalmology
- Vol. 259 (3) , 673-683
Abstract
Purpose To evaluate the effect of visual axis positioning on the optical performance of the Tecnis MIOL and the Diff-aA MIOL. Methods In this prospective, randomized comparative study, 70 eyes of 35 subjects with senile cataract were implanted with the spherical aberration–correcting diffractive, bifocal Tecnis ZLB00 IOL and 60 eyes of 30 age-matched subjects with the spherical aberration neutral, diffractive, bifocal Diffractiva IOL. Observation procedure was performed 1, 3, and 6 months postoperatively. Main outcome measures included uncorrected and corrected distance and near visual acuity, manifest refraction, ocular aberrations, and visual quality metrics with 2 mm and 4 mm pupil and the position of visual axis. Results At the 6-month visit, no significant difference was found in monocular and binocular uncorrected (UDVA) and corrected (CDVA) distance and near (UNVA, CNVA) visual acuity between the groups. Spherical and coma-like aberrations were similar measured with a 2-mm pupil, but with a 4-mm pupil, the SA was significantly larger (in negative direction) in the Diffractiva group. The higher-order Strehl ratio and MTF was significantly larger in the Diffractiva group measured at 2 mm entrance pupil; however, this difference disappeared by the 4-mm pupil measurements. Postoperative angle alpha distance had a significant influence on HO Strehl value. Conclusions The size of angle alpha is a predictive factor of image quality by multifocal IOL patients. Trial Registration Trial registration number and date of registration: NCT04274088, 14.02.2020.Keywords
This publication has 33 references indexed in Scilit:
- Change in Efficiency of Aspheric Intraocular Lenses Based on Pupil DiameterAmerican Journal of Ophthalmology, 2013
- Modulation transfer function and optical quality after bilateral implantation of a +3.00 D versus a +4.00 D multifocal intraocular lensJournal of Cataract & Refractive Surgery, 2012
- Role of angle kappa in patient dissatisfaction with refractive-design multifocal intraocular lensesJournal of Cataract & Refractive Surgery, 2011
- Predictive factor and kappa angle analysis for visual satisfactions in patients with multifocal IOL implantationEye, 2011
- Evaluation of Intraocular Lens Tilt With Anterior Segment Optical Coherence TomographyAmerican Journal of Ophthalmology, 2011
- Optical aberrations and alignment of the eye with ageJournal of Vision, 2010
- Wave aberration of human eyes and new descriptors of image optical quality and visual performanceJournal of Cataract & Refractive Surgery, 2010
- Evaluation of image quality after implantation of 2 diffractive multifocal intraocular lens modelsJournal of Cataract & Refractive Surgery, 2009
- Assessing the optical performance of multifocal (diffractive) intraocular lensesOphthalmic and Physiological Optics, 2008
- Optical performance of monofocal and multifocal intraocular lenses in the human eyeJournal of Cataract & Refractive Surgery, 2008