Multisite Analytical Evaluation of the Abbott ARCHITECT Cyclosporine Assay
- 1 April 2010
- journal article
- research article
- Published by Wolters Kluwer Health in Therapeutic Drug Monitoring
- Vol. 32 (2) , 145-151
Abstract
The objective of this study was to evaluate the analytical performance of the Abbott ARCHITECT Cyclosporine (CsA) immunoassay in 7 clinical laboratories in comparison to liquid chromatography/tandem mass spectrometry (LC/MS/MS), Abbott TDx, Cobas Integra 800, and the Dade Dimension Xpand immunoassay. The ARCHITECT assay uses a whole blood specimen, a pretreatment step with organic reagents to precipitate proteins and extract the drug, followed by a 2-step automated immunoassay with magnetic microparticles coated with anti-CsA antibody and an acridinium-CsA tracer. Imprecision testing at the 7 evaluation sites gave a range of total % coefficient of variations of 7.5%-12.2% at 87.5 ng/mL, 6.6%-14.3% at 411 ng/mL, and 5.2%-10.7% at 916 ng/mL. The lower limit of quantification ranged from 12 to 20 ng/mL. Purified CsA metabolites AM1, AM1c, AM4N, AM9, and AM19 were tested in whole blood by the ARCHITECT assay and showed minimal cross-reactivity at all 7 sites. In particular, AM1 and AM9 cross-reactivity in the ARCHITECT assay, ranged from -2.5% to 0.2% and -0.8% to 2.2%, respectively, and was significantly lower than for the TDx assay, in which the values were 3.2% and 16.1%, respectively. Comparable testing of metabolites in the Dade Dimension Xpand assay at 2 evaluation sites showed cross-reactivity to AM4N (6.4% and 6.8%) and AM9 (2.6% and 3.6%) and testing on the Roche Integra 800 showed cross-reactivity to AM1c (2.4%), AM9 (10.7%), and AM19 (2.8%). Cyclosporine International Proficiency Testing Scheme samples, consisting of both pooled specimens from patients receiving CsA therapy as well as whole-blood specimens supplemented with CsA, were tested by the ARCHITECT assay at 6 sites and showed an average bias of -24 to -58 ng/mL versus LC/MSMS CsA and -2 to -37 ng/mL versus AxSYM CsA. Studies were performed with the ARCHITECT CsA assay on patient specimens with the following results: ARCHITECT CsA assay versus LC/MSMS, average bias of 31 ng/mL; ARCHITECT versus the Dade Dimension assay (4 sites), average biases of -7 to -228 ng/mL; ARCHITECT versus AxSYM and TDx, average biases of -4 and -53 ng/mL, respectively. Spearman correlation coefficients were >or=0.89. The ARCHITECT CsA assay has significantly reduced CsA metabolite interference relative to other immunoassays and is a convenient and sensitive semiautomated method to measure CsA in whole blood.Keywords
This publication has 22 references indexed in Scilit:
- Multi-site analytical evaluation of a chemiluminescent magnetic microparticle immunoassay (CMIA) for sirolimus on the Abbott ARCHITECT analyzerClinical Biochemistry, 2009
- Multi-site Analytical Evaluation of the Abbott ARCHITECT Tacrolimus AssayTherapeutic Drug Monitoring, 2009
- Inter-laboratory differences in sirolimus results from six sirolimus testing centers using HPLC tandem mass spectrometry (LC/MS/MS)Clinica Chimica Acta, 2005
- Therapeutic drug monitoring of cyclosporineTransplantation Proceedings, 2004
- Lack of specificity of cyclosporine immunoassays. Results of a College of American Pathologists Study.Archives of Pathology & Laboratory Medicine, 2003
- Cyclosporin: revisions in monitoring guidelines and review of current analytical methodsAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2002
- Evaluation of a No-Pretreatment Cyclosporin A Assay on the Dade Behring Dimension RxL Clinical Chemistry AnalyzerClinical Chemistry, 2002
- Cyclosporin Measurement – who cares about inaccuracy?Accreditation and Quality Assurance, 2001
- Therapeutic drug monitoring of cyclosporine and tacrolimusClinical Biochemistry, 1998
- A New Biometrical Procedure for Testing the Equality of Measurements from Two Different Analytical Methods. Application of linear regression procedures for method comparison studies in Clinical Chemistry, Part IClinical Chemistry and Laboratory Medicine (CCLM), 1983