Correction for dose-response variations in a scanning liquid ion chamber EPID as a function of linac gantry angle
Authors:
- P W ChinPCP W Chin
- D G LewisDLD G Lewis
- E SpeziESE Spezi
- 23 March 2004
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 49 (8) , N93-N103
Abstract
The response of electronic portal imaging devices (EPIDs) of the scanning liquid ionization chamber (SLIC) type is known to vary with linear accelerator gantry angle. This work considered several contributing factors, quantified the artefacts, monitored their reproducibility and investigated the effects of repeated gantry rotations. Unflatness of up to 5% was found. A correction technique was devised using nonlinear regression of a three-variable sinusoidal modulation. Comparison with two existing techniques found our method to be the most effective, providing a flatness well within 2%. This improved accuracy is expected to benefit more accurate dosimetric studies in particular. The post-acquisition correction process required no change in imaging protocols. Applicability of the new technique was demonstrated on images acquired on different days and with different beam sizes. Since the artefacts compromise both accurate dosimetry and image quality, their successful removal should benefit a broad range of SLIC EPID applications.Keywords
This publication has 17 references indexed in Scilit:
- Monte Carlo simulation of portal dosimetry on a rectilinear voxel geometry: a variable gantry angle solutionPhysics in Medicine & Biology, 2003
- Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapyMedical Physics, 2003
- Multilayer Gafchromic film detectors for breast skin dose determinationin vivoPhysics in Medicine & Biology, 2002
- Clinical use of electronic portal imaging: Report of AAPM Radiation Therapy Committee Task Group 58Medical Physics, 2001
- Radiological thickness measurement using a liquid ionization chamber electronic portal imaging devicePhysics in Medicine & Biology, 1999
- Extraction of primary signal from EPIDs using only forward convolutionMedical Physics, 1997
- Two-dimensional exit dosimetry using a liquid-filled electronic portal imaging device and a convolution modelRadiotherapy and Oncology, 1997
- Calculation of portal dose using the convolution/superposition methodMedical Physics, 1996
- Dosimetric characteristics of a liquid-filled electronic portal imaging deviceInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1995
- A review of electronic portal imaging devices (EPIDs)Medical Physics, 1992