Calibration of a Condensation Particle Counter Using a NIST Traceable Method
Authors:
Open Access
- 5 March 2009
- journal article
- research article
- Published by Taylor & Francis in Aerosol Science and Technology
- Vol. 43 (5) , 425-441
Abstract
This work presents a calibration of a commercial condensation particle counter using National Institute of Standards and Technology (NIST) traceable methods. By the nature of the metrology involved, this work also compares the measurement results of three independent techniques for measuring aerosol concentration: continuous flow condensation particle counter (CPC); aerosol electrometer (AE); and the aerosol concentration derived from microscopic particle counting. Because of the transient nature of aerosol, there are no concentration artifact standards such as exist for particle diameter standards. We employ a mobility classifier to produce a nearly monodisperse, 80 nm, polystyrene latex aerosol. The test aerosol is used as a challenge for the CPC and the AE, and is subsequently filter sampled for electron microscopy. Our test stand design incorporates a continuous CPC aerosol concentration monitor to verify the aerosol stability. The CPC determines particle concentration by single particle counting at a constant sample flow rate. The AE has been calibrated to a NIST traceable current standard. The subsequent aerosol concentration measurement is obtained by determining the electrical current produced by a charged aerosol transported to the detector by a controlled aerosol flow rate. We have NIST traceability for flow rates for all methods and a methodology to calibrate the AE to NIST traceable electrical standards. The latter provides a calibration and a determination of the uncertainty in the aerosol derived current measurement. A bias in the measurements due to multiple charged particles was observed and overcome by using an electrospray aerosol generator to produce the challenge particles. This generator was able to produce aerosol concentrations over the range of 100 particles/cm 3 to 15 000 particles/cm 3 with lower number of dimer particles (≈1%). In our work, independent measurement of aerosol concentration is obtained by quantitatively collecting samples of the airborne polystyrene latex spheres on a small pore filter material and determining the number of particles collected by electron microscopy. Electron micrograph images obtained using a field-emission scanning electron microscope are analyzed using particle counting. We found the relative uncertainty in the aerosol electrometer measurements to be in excess of 100% for particle concentrations of approximately 120 particles/cm 3 and approximately 5% for concentrations above 6000 particles/cm 3 . The uncertainty found by the microscopy method was approximately 3%.Keywords
This publication has 20 references indexed in Scilit:
- CODATA recommended values of the fundamental physical constants: 2002Reviews of Modern Physics, 2005
- Causes of Concentration Differences Between a Scanning Mobility Particle Sizer and a Condensation Particle CounterAerosol Science and Technology, 2003
- Numerical and experimental studies of particle deposition in a tube with a conical contraction—Laminar flow regimeJournal of Aerosol Science, 1995
- Precision, accuracy, uncertainty and traceability and their application to submicrometer dimensional metrologySolid-State Electronics, 1993
- Particle activation and droplet growth processes in condensation nucleus counter—II. Experimental studyJournal of Aerosol Science, 1990
- Dependence of the Performance of TSI 3020 Condensation Nucleus Counter on Pressure, Flow Rate, and TemperatureAerosol Science and Technology, 1990
- Particle activation and droplet growth processes in condensation nucleus counter—I. Theoretical backgroundJournal of Aerosol Science, 1990
- Determining the Size of Non-A, Non-B Hepatitis Virus by FiltrationThe Journal of Infectious Diseases, 1987
- Intercomparison of different “absolute” instruments for measurement of aerosol number concentrationJournal of Aerosol Science, 1982
- Intercomparison of optical particle counters under conditions of normal operationAtmospheric Environment (1967), 1980