A systematic MEMS sensor calibration framework
Authors:
Open Access
- 27 February 2015
- journal article
- research article
- Published by Copernicus GmbH in Journal of Sensors and Sensor Systems
- Vol. 4 (1) , 97-102
Abstract
In this paper we present a systematic method to determine sets of close to optimal sensor calibration points for a polynomial approximation. For each set of calibration points a polynomial is used to fit the nonlinear sensor response to the calibration reference. The polynomial parameters are calculated using ordinary least square fit. To determine the quality of each calibration, reference sensor data is measured at discrete test conditions. As an error indicator for the quality of a calibration the root mean square deviation between the calibration polynomial and the reference measurement is calculated. The calibration polynomials and the error indicators are calculated for all possible calibration point sets. To find close to optimal calibration point sets, the worst 99% of the calibration options are dismissed. This results in a multi-dimensional probability distribution of the probably best calibration point sets. In an experiment, barometric MEMS (micro-electromechanical systems) pressure sensors are calibrated using the proposed calibration method at several temperatures and pressures. The framework is applied to a batch of six of each of the following sensor types: Bosch BMP085, Bosch BMP180, and EPCOS T5400. Results indicate which set of calibration points should be chosen to achieve good calibration results.Keywords
This publication has 13 references indexed in Scilit:
- A Framework for Calibration of Barometric MEMS Pressure SensorsProcedia Engineering, 2014
- Flip-chip packaging of piezoresistive barometric pressure sensors2013
- A tutorial on support vector regressionStatistics and Computing, 2004
- Kalibrierung von integrierten Drucksensoren im Waferverbund (On-Wafer Calibration for Pressure Sensors)TM - Technisches Messen, 2003
- A non-iterative, polynomial, 2-dimensional calibration method implemented in a microcontroller2002
- Evaluation of measurement uncertainty for analytical procedures using a linear calibration functionAccreditation and Quality Assurance, 2002
- Calibration uncertaintyAccreditation and Quality Assurance, 2002
- Digital compensation of high-performance silicon pressure transducersSensors and Actuators A: Physical, 1990
- Digital compensation of sensorsJournal of Physics E: Scientific Instruments, 1987
- A general digital linearising method for transducersJournal of Physics E: Scientific Instruments, 1985