Including the effect of motion artifacts in noise and performance analysis of dual-energy contrast-enhanced mammography
- 30 November 2012
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 57 (24) , 8405-8425
Abstract
In contrast-enhanced mammography (CEM), the dual-energy dual-exposure technique, which can leverage existing conventional mammography infrastructure, relies on acquiring the low- and high-energy images using two separate exposures. The finite time between image acquisition leads to motion artifacts in the combined image. Motion artifacts can lead to greater anatomical noise in the combined image due to increased mismatch of the background tissue in the images to be combined, however the impact has not yet been quantified. In this study we investigate a method to include motion artifacts in the dual-energy noise and performance analysis. The motion artifacts are included via an extended cascaded systems model. To validate the model, noise power spectra of a previous dual-energy clinical study are compared to that of the model. The ideal observer detectability is used to quantify the effect of motion artifacts on tumor detectability. It was found that the detectability can be significantly degraded when motion is present (e.g., detectability of 2.5 mm radius tumor decreased by approximately a factor of 2 for translation motion on the order of 1000 μm). The method presented may be used for a more comprehensive theoretical noise and performance analysis and fairer theoretical performance comparison between dual-exposure techniques, where motion artifacts are present, and single-exposure techniques, where low- and high-energy images are acquired simultaneously and motion artifacts are absent.Keywords
This publication has 54 references indexed in Scilit:
- K-Edge Imaging Using Dual-Layer and Single-Layer Large Area Flat Panel ImagersIEEE Transactions on Nuclear Science, 2012
- Single-layer and dual-layer contrast-enhanced mammography using amorphous selenium flat panel detectorsPhysics in Medicine & Biology, 2011
- Analysis of multilayer and single layer X-ray detectors for contrast-enhanced mammography using imaging task2011
- Multilayer x-ray detector for contrast-enhanced digital subtraction mammography2010
- Comparison of dual energy detector system performanceMedical Physics, 2004
- Megalopinakophobia: its symptoms and cures2001
- Medical image restoration of dynamic lungs using optical transfer function of lung motionJournal of Biomedical Optics, 2001
- Linear response theory for detectors consisting of discrete arraysMedical Physics, 2000
- Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammographyMedical Physics, 1997
- Objective assessment of image quality: effects of quantum noise and object variabilityJournal of the Optical Society of America A, 1990