The erlanger silver catheter:In vitro results for antimicrobial activity
Authors:
- 1 January 1999
- journal article
- research article
- Published by Springer Nature in Infection
- Vol. 27 (S1) , S24-S29
Abstract
The antimicrobial activity of a silver-impregnated polymer catheter (the Erlanger silver catheter) was demonstrated by determining the microbial adhesion to the surface of the catheter and by measuring the rate of proliferation (viability) of microorganisms at this site. On the surface of a catheter impregnated with silver, according to previously described methods, the bacterial adhesion ofStaphylococcus epidermidis is reduced by 28–40%. Bacterial proliferation on the surface of the catheter and biofilm production are also substantially reduced by the elution of free silver ions from the catheter matrix. Bacteriostatic and bactericidal activities can be determined. The antimicrobial efficacy of the silver catheter is not reduced by blood components. There is no loss in antimicrobial activity for weeks after preincubation in water or phosphate buffered saline. The antimicrobial activity depends on the extent of the active silver surface.Keywords
This publication has 7 references indexed in Scilit:
- Multiple parameters for the comprehensive evaluation of the susceptibility of Escherichia coli to the silver ionBioMetals, 1998
- Characterization of Tn917 insertion mutants of Staphylococcus epidermidis affected in biofilm formationInfection and Immunity, 1996
- Antimicrobial activity and biocompatibility of polyurethane and silicone catheters containing low concentrations of Silver: A new perspective in prevention of polymer-associated foreign-body-infectionsZentralblatt für Bakteriologie, 1995
- In vitro Evaluation of the Antimicrobial Efficacy and Biocompatibility of a Silver-Coated Central Venous CatheterJournal of Biomaterials Applications, 1994
- Surface Antimicrobial Activity of Heparin-Bonded and Antiseptic-Impregnated Vascular CathetersThe Journal of Infectious Diseases, 1993
- Parallel induction by glucose of adherence and a polysaccharide antigen specific for plastic-adherent Staphylococcus epidermidis: evidence for functional relation to intercellular adhesionInfection and Immunity, 1992
- Host Factors Selectively Increase Staphylococcal Adherence on Inserted Catheters: A Role for Fibronectin and Fibrinogen or FibrinThe Journal of Infectious Diseases, 1989