Validation of temporal gait metrics from three IMU locations to the gold standard force plate
- 18 October 2016
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 2016, 667-671
Abstract
The purpose of this work is to compare temporal gait parameters from three different IMU locations to the gold standard force platform. 33 subjects (12 F, 21 M) performed twenty gait trials each while wearing inertial measurement units (IMUs) on the trunk, both shanks and both feet. Data was simultaneously collected from a laboratory embedded force plate. Step times were derived from the raw IMU data at the three IMU locations using methods that have been shown to be accurate. Step times from all locations were valid compared to the force plate. Foot IMU step time was the most accurate (Pearson = .991, CI width = 3.00e2), the trunk IMU was the next most accurate (Pearson = .974, CI width = 4.85e2) and shank step time was the least accurate (Pearson = .958, CI width = 6.80e2). All three sensing locations result in valid estimations of step time compared to the gold standard force plate. These results suggest that the foot location would be most appropriate for clinical applications where very precise temporal parameter detection is required.Keywords
This publication has 33 references indexed in Scilit:
- iGAIT: An interactive accelerometer based gait analysis systemComputer Methods and Programs in Biomedicine, 2012
- Toward Automated, At-Home Assessment of Mobility Among Patients With Parkinson Disease, Using a Body-Worn AccelerometerNeurorehabilitation and Neural Repair, 2011
- Real-time gait event detection for normal subjects from lower trunk accelerationsGait & Posture, 2010
- Concurrent validity of a trunk tri-axial accelerometer system for gait analysis in older adultsGait & Posture, 2009
- Accelerometry: A technique for quantifying movement patterns during walkingGait & Posture, 2008
- Gait event detection using linear accelerometers or angular velocity transducers in able-bodied and spinal-cord injured individualsGait & Posture, 2006
- Gait Assessment in Parkinson's Disease: Toward an Ambulatory System for Long-Term MonitoringIEEE Transactions on Biomedical Engineering, 2004
- Three dimensional inertial sensing of foot movements for automatic tuning of a two-channel implantable drop-foot stimulatorMedical Engineering & Physics, 2002
- Thirteen Ways to Look at the Correlation CoefficientThe American Statistician, 1988
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986