• 1 January 1997
    • journal article
    • Vol. 27  (1) , 36-42
    • PMID:9000583
Abstract
We report on preliminary investigations of the use of an image analysis system to perform preliminary algorithmic classification of images of fluorochrome-labeled cells followed by capture of gray-level images of potentially abnormal cells for analysis by a neural network. Cells were labeled with an antibody against a bladder cancer tumor-associated antigen, and the neural net was used to distinguish true-positive cells from negative cells, false-positive cells (autofluorescent or nonspecific labeling), and cell-sized artifacts. Gray-level cell images were digitized and processed for analysis by a feed-forward neural network using back-propagation. The network was trained and tested with two independent image sets. Various network configurations and activation functions were investigated, including a sinusoidal activation function. At high power, the network agreed completely with the human observer's classification. At low power, a strong clustering of cells classified by the network with expert classification was seen, while the neural network showed roughly 75% concordance with the human observer. In addition, a set of four features extracted from raw cell images were investigated. The features were: shape factor, texture, area, and average pixel intensity. A network trained with these features performed better than one operating with gray-level images. We conclude that using neural networks to recognize and classify images captured by an image analysis microscope is feasible.