Comparison of the rotational stability between plate-haptic toric and C-loop haptic toric IOLs in myopic eyes
Authors:
- Jiaqi MengJMJiaqi Meng
- Wenwen HeWHWenwen He
- Xianfang RongXRXianfang Rong
- Yi LuYLYi Lu
- 1 October 2020
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 46 (10) , 1353-1359
Abstract
Purpose: To compare the rotational stability of a plate-haptic toric intraocular lens (IOL) vs a C-loop haptic toric IOL in myopic cataract eyes. Setting: Eye and Ear, Nose, and Throat Hospital of Fudan University, China. Design: Prospective, randomized, controlled study. Methods: Cataract eyes with axial length (AL) more than 24.5 mm were randomly assigned to receive implantation of a C-loop haptic toric IOL (AcrySof Toric IOL) (Group A) or a plate-haptic toric IOL (AT TORBI 709M IOL) (Group B). IOL rotation, residual astigmatism, visual acuity, and higher-order aberrations (HOAs) evaluated with OPD-Scan III aberrometer were compared at 3 months postoperatively. Results: In total, 62 eyes of 62 patients were eligible for analysis: 31 in Group A and 31 in Group B. The mean rotation of toric IOLs was greater in Group A than that in Group B (8.00 ± 3.60 degrees vs 4.42 ± 3.24 degrees, respectively, P < .001), especially when IOLs were vertically placed. IOL rotation was positively correlated with AL in Group A, whereas no such correlations were found in Group B. Residual astigmatism in Group A was greater than that in Group B (−0.76 ± 0.30 diopter [D] vs −0.51 ± 0.29 D, respectively, P = .001). Fewer eyes achieved residual astigmatism of 0.50 D or less in Group A than in Group B (38.71% vs 64.52%). Group A had worse postoperative uncorrected visual acuity and higher total HOAs and coma for a 6.0 mm pupil than Group B, whereas postoperative corrected visual acuity was not different between the 2 groups. Conclusions: The plate-haptic toric IOL might be a better choice for myopic cataract eyes with corneal astigmatism because of reduced postoperative rotation.Keywords
This publication has 24 references indexed in Scilit:
- Comparative evaluation of two toric intraocular lenses for correcting astigmatism in patients undergoing phacoemulsificationIndian Journal Of Ophthalmology, 2018
- Relationship between age, corneal astigmatism, and ocular dimensions with reference to astigmatism in eyes undergoing routine cataract surgeryEye, 2016
- Factors influencing 1-year rotational stability of AcrySof Toric intraocular lensesBritish Journal of Ophthalmology, 2015
- Clinical outcomes and rotational stability of a 4-haptic toric intraocular lens in myopic eyesJournal of Cataract & Refractive Surgery, 2014
- Residual astigmatism produced by toric intraocular lens rotationJournal of Cataract & Refractive Surgery, 2011
- Analysis of biometry and prevalence data for corneal astigmatism in 23 239 eyesJournal of Cataract & Refractive Surgery, 2010
- Evaluating the eye's rotational stability during standard photographyJournal of Cataract & Refractive Surgery, 2005
- Astigmatism and toric intraocular lensesCurrent Opinion in Ophthalmology, 2000
- Postoperative intraocular lens rotationOphthalmology, 1999
- Biometry of 7,500 Cataractous EyesAmerican Journal of Ophthalmology, 1980