On the characterization and uncertainty analysis of radiochromic film dosimetry
Authors:
- 5 May 2009
- journal article
- radiation measurement-physics
- Published by Wiley in Medical Physics
- Vol. 36 (6Part1) , 1931-1946
Abstract
Radiochromic film is a dosimeter of choice in applications requiring high spatial resolution, two dimensional measurements, or minimum perturbation of the beam fluence. Since the measurement uncertainty in Gafchromic film dosimetry is thought to be significant compared to that of ionization chambers, a rigorous method to evaluate measurement uncertainties is desired. This article provides a method that takes into account the correlation between fit parameters as well as single dose values in order to obtain accurate uncertainties in absolute and relative measurements. A complete portrait of all sources of uncertainty in Gafchromic film dosimetry is given. The parametrization of variance as a function of the number of averaged pixels is obtained in order to accurately predict the uncertainty as a function of the size of the region of interest. The choice of functional form for the sensitometric curve is based on four criteria and a convergence of global net optical density uncertainty to 0.0013 is demonstrated. A minimum number of 12 points is recommended to characterize the sensitometric curve to a sufficient precision on the uncertainty estimation. Uncertainty levels of 0.9% on absolute dose measurements and 0.45% on relative measurements are achieved using a 12‐point calibration curve with 220 cGy and repeating measurements five times. Uncertainties of 0.8% and 0.4% are achievable when using 35 points during film characterization. Ignoring covariance terms is shown to lead to errors in the estimation of uncertainty.Keywords
Funding Information
- Natural Sciences and Engineering Research Council of Canada (344218)
This publication has 14 references indexed in Scilit:
- Analysis of uncertainties in Gafchromic® EBT film dosimetry of photon beamsPhysics in Medicine & Biology, 2008
- GafChromic EBT film dosimetry with flatbed CCD scanner: A novel background correction method and full dose uncertainty analysisMedical Physics, 2008
- Accurate dosimetry with GafChromic™ EBT film of a 6MV photon beam in water: What level is achievable?Medical Physics, 2008
- Important considerations for radiochromic film dosimetry with flatbed CCD scanners and EBT filmMedical Physics, 2006
- Precise radiochromic film dosimetry using a flat-bed document scannerMedical Physics, 2005
- Ionization chamber‐based reference dosimetry of intensity modulated radiation beamsMedical Physics, 2004
- Dosimetric properties of improved GafChromic films for seven different digitizersMedical Physics, 2004
- Characteristics of sensitometric curves of radiographic filmsMedical Physics, 2003
- Physical aspects of dynamic stereotactic radiosurgery with very small photon beams (1.5 and 3 mm in diameter)Medical Physics, 2003
- Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55Medical Physics, 1998