All-solid-state Lead(II)–Selective Potentiometric Sensor Based on 9,10-dibromo-2-(propylthio)anthracene-1,4-dione
Open Access
- 31 March 2025
- journal article
- Published by Canakkale Onsekiz Mart University in Journal of Advanced Research in Natural and Applied Sciences
- Vol. 11 (1) , 53-62
Abstract
Ion-selective sensors are widely used in routine analysis of ionic species. In this research, a highly selective potentiometric sensor was developed, which is sensitive against Pb2+ ions compared to other common inorganic cations. The optimum composition of the developed sensor was determined to be as 3.0% ionophore, 64.0% bis(2–ethylhexyl)adipate (DEHA), 32.0% poly (vinyl chloride) (PVC), and 1.0% anion excluder (KTpClPB). The Pb2+–selective sensor had a Nernstian response of 28.7±2.2 mV/decade over a wide linear range of 1.0×10-5–1.0×10-1 M and a low detection limit of 2.77×10-6 M. The newly developed sensor worked over a wide pH range (4.0–10.0), and the response time was short (7s). The sensor, which had good reproducibility, was applied to various water samples and determined Pb2+ ions with very high recoveries.Keywords
This publication has 27 references indexed in Scilit:
- Protic Ionic Liquid Based Potentiometric Sensors: High Selectivity Detection of Silver(I) IonsChemistrySelect, 2025
- Synthesis, characterization and sensor properties of a new sulfonyl hydrazone derivative molecule: potentiometric determination of Pb(II) ionsChemical Papers, 2024
- Surface characterization and electroanalytical applications of the newly developed copper(II)-selective potentiometric sensorAnalytical Sciences, 2023
- Molecular mechanisms of lead neurotoxicity2021
- Assessing the mechanism of DNA damage induced by lead through direct and indirect interactionsJournal of Photochemistry and Photobiology B: Biology, 2014
- Ion‐Selective Electrodes With Ionophore‐Doped Sensing Membranes2012
- Central nervous system cytokine gene expression: Modulation by leadJournal of Biochemical and Molecular Toxicology, 2011
- Effect of lead toxicity on coenzyme Q levels in rat tissuesFood and Chemical Toxicology, 2010
- Lead induces oxidative stress, DNA damage and alteration of p53, Bax and Bcl-2 expressions in miceFood and Chemical Toxicology, 2008
- All-Solid-State PVC Membrane Ag + -Selective Electrodes Based on Diaza-18-Crown-6 CompoundsMicrochimica Acta, 2004