Application of failure mode and effects analysis to optimization of linac quality controls protocol
- 11 April 2019
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 46 (6) , 2541-2555
Abstract
Purpose: To apply Failure Mode and Effects Analysis (FMEA) to optimize linac quality control (QC) protocol in order to ensure patient safety and treatment quality, taking maximum advantage of the available resources.Material and methods: Each parameter tested by the QC was considered as a potential failure mode (FM). For each FM, likelihood of occurrence (O), severity of effect (S), and lack of detectability (D) were evaluated and corresponding Risk Priority Number (RPN) was calculated from the product of three indexes. The scores were assigned using two methods: (a) A survey submitted to the medical physicists; (b) A semi‐quantitative analysis (SQA) performed through: simulation of FMs in the treatment planning system; studies reported in literature; results obtained by the QC data analysis. A weighted RPN for all FMs was calculated taking into account both the methods. For each linac, the tests were then sorted by their frequency and the RPN value.Results: A high variability was found in the scores of the survey, although in many it was reduced in RPN values, highlighting the more relevant tests as on beam output and imaging system. Integrating these results with those obtained by SQA, the RPN‐based ranking of tests has been provided considering the specific use of the accelerator: for example, more accurate tests on dose modulation and multileaf collimator speed were required in linacs where intensity‐modulated treatment is performed, while, more specific tests on couch and jaw position indicators were necessary where treatments with multiple isocenters and/or junctions between adjacent fields were often delivered.Conclusions: Failure Mode and Effects Analysis is a useful tool to prioritize the linac QCs, taking into account the specific equipment and clinical practice. The integration of SQA and survey results reduces subjectivity of the FMEA scoring and allows to optimize linac QCs without “losing” the expertise and experience of medical physicists and clinical staff.Keywords
This publication has 30 references indexed in Scilit:
- AAPM Medical Physics Practice Guideline 8.a.: Linear accelerator performance testsJournal of Applied Clinical Medical Physics, 2017
- The report of Task Group 100 of the AAPM: Application of risk analysis methods to radiation therapy quality managementMedical Physics, 2016
- A streamlined failure mode and effects analysisMedical Physics, 2014
- Application of failure mode and effects analysis to treatment planning in scanned proton beam radiotherapyRadiation Oncology, 2013
- Application of Failure Mode and Effects Analysis to Intraoperative Radiation Therapy Using Mobile Electron Linear AcceleratorsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012
- Failure mode and effect analysis‐based quality assurance for dynamic MLC tracking systemsMedical Physics, 2010
- Task Group 142 report: Quality assurance of medical acceleratorsa)Medical Physics, 2009
- Evaluation of Safety in a Radiation Oncology Setting Using Failure Mode and Effects AnalysisInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Quality Assurance of Radiotherapy in Cancer Treatment: Toward Improvement of Patient Safety and Quality of CareJapanese Journal of Clinical Oncology, 2008
- A Method for Evaluating Quality Assurance Needs in Radiation TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008