Transcatheter tricuspid valve‐in‐valve in patients with transvalvular device leads

Authors:

Abstract
Objective: To describe the effects of percutaneously implanted valve‐in‐valve in the tricuspid position for patients with pre‐existing transvalvular device leads.Methods: In this case series, we describe implantation of the Melody valve and SAPIEN XT valve within dysfunctional bioprosthetic tricuspid valves in three patients with transvalvular device leads.Results: In all cases, the valve was successfully deployed and device lead function remained unchanged. In 1/3 cases with 6‐month follow‐up, device lead parameters remain unchanged and transcatheter valve‐in‐valve function remains satisfactory.Conclusions: Transcatheter tricuspid valve‐in‐valve is feasible in patients with pre‐existing transvalvular devices leads. Further study is required to determine the long‐term clinical implications of this treatment approach.