Quality control of spirometry: a lesson from the BRONCUS trial

Authors:

Abstract
This report describes the quality control programme used within the Bronchitis Randomized on N-acetylcysteine (NAC) Cost-Utility Study, a trial designed to assess the decline in lung function, exacerbation rate, health status, and cost-effectiveness with NAC or a placebo in 523 patients with chronic obstructive pulmonary disease over a 3-yr period.Spirometry was scored from 0 (worst quality) to 6 (best quality). The mean score of 314 spirometries from 243 patients evaluated during the trial was 5.63±0.83. Linear regression analysis of the scores of 47 participating centres plotted against the time at which spirometries were performed yielded an intercept of 5.7±0.5 and a slope of -0.0001±0.001, which suggests that the initial high quality was maintained over time.Retrospective examination of a further 345 postbronchodilator spirometries from 208 patients with a forced expiratory volume at one second exceeding the mean individual value recorded over the study in excess of 20% revealed a slightly lower quality of the start-of-test manoeuvre compared with the 314 spirometries.In conclusion, these findings would suggest that the quality control programme is likely to have helped achieve and maintain long-term spirometry performance in the Bronchitis Randomized on N-acetylcysteine (NAC) Cost-Utility Study trial. Special care should be paid to the spirometries whose forced expiratory volume in one second values exceed the mean value.

This publication has 13 references indexed in Scilit: