Abstract
This report describes the quality control programme used within the Bronchitis Randomized on N-acetylcysteine (NAC) Cost-Utility Study, a trial designed to assess the decline in lung function, exacerbation rate, health status, and cost-effectiveness with NAC or a placebo in 523 patients with chronic obstructive pulmonary disease over a 3-yr period.Spirometry was scored from 0 (worst quality) to 6 (best quality). The mean score of 314 spirometries from 243 patients evaluated during the trial was 5.63±0.83. Linear regression analysis of the scores of 47 participating centres plotted against the time at which spirometries were performed yielded an intercept of 5.7±0.5 and a slope of -0.0001±0.001, which suggests that the initial high quality was maintained over time.Retrospective examination of a further 345 postbronchodilator spirometries from 208 patients with a forced expiratory volume at one second exceeding the mean individual value recorded over the study in excess of 20% revealed a slightly lower quality of the start-of-test manoeuvre compared with the 314 spirometries.In conclusion, these findings would suggest that the quality control programme is likely to have helped achieve and maintain long-term spirometry performance in the Bronchitis Randomized on N-acetylcysteine (NAC) Cost-Utility Study trial. Special care should be paid to the spirometries whose forced expiratory volume in one second values exceed the mean value.Keywords
This publication has 13 references indexed in Scilit:
- Effects of N-acetylcysteine on outcomes in chronic obstructive pulmonary disease (Bronchitis Randomized on NAC Cost-Utility Study, BRONCUS): a randomised placebo-controlled trialThe Lancet, 2005
- Validity of spirometric testing in a general practice population of patients with chronic obstructive pulmonary disease (COPD)Thorax, 2003
- Quality Control of Spirometry in the Elderly: The SA.R.A. StudyAmerican Journal of Respiratory and Critical Care Medicine, 2000
- Volume History Effects and Airway Responsiveness in Middle-Aged and Older Men. The Normative Aging StudyAmerican Journal of Respiratory and Critical Care Medicine, 1997
- Standardization of Spirometry, 1994 Update. American Thoracic SocietyAmerican Journal of Respiratory and Critical Care Medicine, 1995
- Variability of FVC and FEV1 due to technician, team, device and subject in an eight centre study: three quality control studies in SAPALDIA. Swiss Study on Air Pollution and Lung Disease in AdultsEuropean Respiratory Journal, 1995
- Dependence of maximal flow-volume curves on time course of preceding inspirationJournal of Applied Physiology, 1993
- Lung volumes and forced ventilatory flowsEuropean Respiratory Journal, 1993
- Spirometry in the Lung Health Study: 1. Methods and Quality ControlAmerican Review of Respiratory Disease, 1991
- Effect of Effort on Measurement of Forced Expiratory Volume in One SecondAmerican Review of Respiratory Disease, 1987