A Bioanalytical Method for Quantification of Telmisartan in rat Plasma; Development, Validation and Application to Pharmacokinetic Study

Authors:

Abstract
The telmisartan was determined in a rat plasma using developed and validated a reversed-phase high performance liquid chromatographic (HPLC). The pre-treatment of the plasma sample involving liquid-liquid extraction using ethanol as the extracting solvent. The HPLC method validation has been shown a linear calibration curve over a plasma concentrations range of 0.7 to 10µg/mL with a correlation coefficient of 0.9979, the limit of detection and the limit of quantification were determined to be 0.025µg/ml and 0.07µg/ml, respectively. The precision and accuracy were in an acceptable limit. The pharmacokinetic parameters of telmisartan were adequately evaluated following a single oral dose (4mg/kg) in Sprague-Dawley rats. The results observed conclude that the developed bioanalytical HPLC method is appropriate and applicable as an analytical tool in the pharmacokinetic study of telmisartan.

This publication has 14 references indexed in Scilit: