SNR versus resolution in 3D 1H MRS of the human brain at high magnetic fields

Authors:

Abstract
It is commonly accepted that the signal‐to‐noise ratio (SNR = peak‐signal/RMS‐noise) per‐unit‐time of proton MR spectroscopy (1H‐MRS) is linearly proportional to the voxel volume. Consequently, with a headcoil and 30‐min acquisition, 1 cm3 is considered the SNR‐limited spatial resolution barrier in the human brain. However, since local linewidths, Δυ* = (πT)−1, at high magnetic fields (B0), are dominated by regional inhomogeneities (ΔB0), i.e., T ≪ T2, reducing the voxel dimensions may increase T. This could compensate, in part, for signal loss with volume decrease. It is shown that for two cubic voxels of sides l1 and l2, l1 > l2, as the volume decreases by (l1/l2)3, their SNR ratio is reduced by only (l1/l2)2 due to a commensurate T increase of l1/l2. This is demonstrated in a phantom and the brains of volunteers, with 3D 1H‐MRS in a headcoil at 4 T. It is shown that while the cubic voxels' dimensions were all halved, reducing their volume eightfold, their metabolites' SNR decreased only fourfold, due to their Δυ*s' twofold decrease. In other words, both spatial and spectral resolutions were doubled at a significantly, ×2, smaller‐than‐expected SNR loss. This advantage was exploited to produce quality high spatial resolution, 0.75 × 0.75 × 0.75 cm3, metabolic maps in a 27‐min acquisition. Magn Reson Med 46:1049–1053, 2001.