Histological Evaluation of the Serf EVL Evolution Implant: A Pilot Study in a Dog Model
Authors:
- Stuart FroumSFStuart Froum
- Dennis TarnowDTDennis Tarnow
- Ziad JalboutZJZiad Jalbout
- Jean-Pierre BrunJBJean-Pierre Brun
- Robert FromentalRFRobert Fromental
- 1 March 2003
- journal article
- research article
- Published by Wolters Kluwer Health in Implant Dentistry
- Vol. 12 (1) , 69-74
Abstract
The purpose of this study was to compare bone-to-implant contact (BIC) of the one-stage sandblasted surface EVL implant (SERF, Decines, France) with that of the two-stage turned (machined) Branemark Nobel BioCare implant in a dog model system. Three 1-year old female beagle dogs (canine) were acquired, examined, and quarantined. In dog no. 1, three test implants were placed on one side of the jaw, and four control implants were placed on the other side. In dog no. 2, two test implants were placed in alternation with two control implants on the left and right sides (total 4 tests, 4 controls). In dog no. 3, three test implants were placed on one side and three control implants on the other. After a 3-month period of healing, the dogs were killed and the jaws sectioned for histological preparation. Eleven implants, six tests, and five controls were available for histological evaluation. The percent of BIC for the test implants varied from 24.9% to 61.6%, with an average of 42.7%. The percent of BIC for the control implants varied from 22.1% to 42.6%, with an average of 27.4%. The degree of BIC of the EVL implants was on average greater than that of the turned surface of the Branemark implant. However, because of the limited number of implants placed in this pilot study, the results should be interpreted with caution.Keywords
This publication has 12 references indexed in Scilit:
- Interface shear strength of titanium implants with a sandblasted and acid-etched surface: A biomechanical study in the maxilla of miniature pigsJournal of Biomedical Materials Research, 1999
- Long‐term evaluation of non‐submerged ITI implants. Part 1: 8‐year life table analysis of a prospective multi‐center study with 2359 implants.Clinical Oral Implants Research, 1997
- Retrospective multicenter analysis of 3i endosseous dental implants placed over a five‐year periodClinical Oral Implants Research, 1996
- Torque and histomorphometric evaluation of c.p. titanium screws blasted with 25- and 75-?m-sized particles of Al2O3Journal of Biomedical Materials Research, 1996
- Experimental study of turned and grit-blasted screw-shaped implants with special emphasis on effects of blasting material and surface topographyBiomaterials, 1996
- A histomorghometric study of screw‐shaped and removal torque titanium implants with three different surface topographiesClinical Oral Implants Research, 1995
- The cutting-grinding technique for histologic preparation of undecalcified bone and bone-anchored implantsOral Surgery, Oral Medicine, Oral Pathology, 1992
- A multicenter report on osseointegrated oral implantsThe Journal of Prosthetic Dentistry, 1988
- A method for the study of undecalcified bones and teeth with attached soft tissues*Journal of Oral Pathology & Medicine, 1982
- A 15-year study of osseointegrated implants in the treatment of the edentulous jawInternational Journal of Oral Surgery, 1981