Interferences with Two-site Immunoassays by Human Anti-mouse Antibodies Formed by Patients Treated with Monoclonal Antibodies: Comparison of Different Blocking Reagents
Author:
Open Access
- 1 August 1998
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 44 (8) , 1742-1744
Abstract
The most widely used approach for reducing interferences with two-site immunoassays by human anti-mouse antibodies (HAMAs) developed by many patients after exposure to murine immunoglobulins is to include high amounts of nonspecific mouse IgG within the assay buffer (1)(2). Recently, Mössner et al. (3) have developed a polymerized form of murine IgG (MAK33), which they found to be superior to normal mouse IgG for blocking HAMA interferences. In contrast, we observed false-positive values for the cancer antigen 125 (CA 125), because of HAMA interferences in samples from ovarian cancer patients treated with the anti-CA-125 antibody OC125, which could be corrected by preincubation with mouse IgG but not with MAK33 (4). The aim of the present study was to examine whether treatment with other monoclonal antibodies gives rise to additional HAMAs that are insensitive to MAK33 and to further characterize this specific HAMA response.Keywords
This publication has 10 references indexed in Scilit:
- Immunological Responses to the Tumor-Associated Antigen CA15 in Patients with Advanced Ovarian Cancer Induced by the Murine Monoclonal Anti-idiotype Vaccine ACA125Hybridoma, 1997
- Different efficacy of various blocking reagents to eliminate interferences by human antimouse antibodies with a two-site immunoassayClinical Biochemistry, 1996
- A Monoclonal Antiidiotypic Antibody ACA 125 Mimicking the Tumor-Associated Antigen CA 125 for Immunotherapy of Ovarian CancerHybridoma, 1995
- Antibody-Targeted Photodynamic TherapyHybridoma, 1993
- Eliminating Interference from Heterophilic Antibodies in a Two-Site Immunoassay for Creatine Kinase Mb by Using F(ab')2 Conjugate and Polyclonal Mouse IgGClinical Chemistry, 1992
- "Sandwich"-type immunoassay of carcinoembryonic antigen in patients receiving murine monoclonal antibodies for diagnosis and therapy.Clinical Chemistry, 1988
- Heterophilic antibodies: a problem for all immunoassays.Clinical Chemistry, 1988
- Human immune response to multiple injections of murine monoclonal IgG.The Journal of Immunology, 1985
- Reactivity of a monoclonal antibody with human ovarian carcinoma.Journal of Clinical Investigation, 1981
- A spectrum of monoclonal antibodies reactive with human mammary tumor cells.Proceedings of the National Academy of Sciences, 1981