Formulation of objective indices to quantify machine failure risk analysis for interruptions in radiotherapy
Open Access
- 16 December 2020
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 22 (1) , 165-173
Abstract
Objectives: To evaluate the effect of interruption in radiotherapy due to machine failure in patients and medical institutions using machine failure risk analysis (MFRA).Material and methods: The risk of machine failure during treatment is assigned to three scores (biological effect, B; occurrence, O; and cost of labor and repair parts, C) for each type of machine failure. The biological patient risk (BPR) and the economic institution risk (EIR) are calculated as the product of and () and and (), respectively. The MFRA is performed in two linear accelerators (linacs).Result: The multileaf collimator (MLC) fault has the highest BPR and second highest EIR. In particular, TrueBeam has a higher BPR and EIR for MLC failures. The total EIR in TrueBeam was significantly higher than that in Clinac iX. The minor interlock had the second highest BPR, whereas a smaller EIR. Meanwhile, the EIR for the LaserGuard fault was the highest, and that for the monitor chamber fault was the second highest. These machine failures occurred in TrueBeam. The BPR and EIR should be evaluated for each linac. Further, the sensitivity of the BPR, it decreased with higher and α/β values. No relative difference is observed in the BPR for each machine failure when and α/β were varied.Conclusion: The risk faced by patients and institutions in machine failure may be reduced using MFRA.Advances in knowledge: For clinical radiotherapy, interruption can occur from unscheduled downtime with machine failures. Interruption causes sublethal damage repair. The current study evaluated the effect of interruption in radiotherapy owing to machine failure on patients and medical institutions using a new method, that is, machine failure risk analysis.Keywords
This publication has 20 references indexed in Scilit:
- The report of Task Group 100 of the AAPM: Application of risk analysis methods to radiation therapy quality managementMedical Physics, 2016
- A model for preemptive maintenance of medical linear accelerators—predictive maintenanceRadiation Oncology, 2016
- A Generalized Linear-Quadratic Model for Radiosurgery, Stereotactic Body Radiation Therapy, and High–Dose Rate BrachytherapyScience Translational Medicine, 2010
- Evaluation of Safety in a Radiation Oncology Setting Using Failure Mode and Effects AnalysisInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- The Linear-Quadratic Model Is an Appropriate Methodology for Determining Isoeffective Doses at Large Doses Per FractionSeminars in Radiation Oncology, 2008
- Some implications of linear‐quadratic‐linear radiation dose‐response with regard to hypofractionationMedical Physics, 2008
- The role of radiotherapy in cancer treatmentCancer, 2005
- Conditions for the equivalence of continuous to pulsed low dose rate brachytherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1991
- Dose fractionation, dose rate and iso-effect relationships for normal tissue responsesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1982
- The mechanism of the induction by radiation of chromosome aberrations inTradescantiaJournal of Genetics, 1942