Density Changes in Thermometer Glasses

Authors:

Abstract
A study was made of the density changes in the thermometer glasses, Corning 0041, 7560, 1720 and an experimental glass, Corning EXP175EK. Thermometers fabricated from these glasses were given similar heat‐treatments in their respective annealing ranges. Measurements were then made of the ice point changes at various fixed temperatures from room temperature to the normal upper useful limits of the glasses. Holding periods extended to 1 year. The ice point change in a thermometer is a measure of the per cent density change in the glass from which the bulb is fabricated. The relation of this change in density to temperature for a given duration of roasting appears to be well represented by two curves of the form of a constant times a temperature to a power. The value of the power appears to change rather abruptly at some temperature, T 0 , in the neighborhood of 150°C. below the strain point temperature and to be constant in the regions on either side of T 0 . At equal temperatures below the strain point the behavior of the glasses is quite similar.