Electrocardiogram Devices for Home Use: Technological and Clinical Scoping Review
Authors:
- Alejandra Zepeda-EchavarriaAZAlejandra Zepeda-Echavarria
- Rutger R van de LeurRLRutger R van de Leur
- Meike van SleuwenMSMeike van Sleuwen
- Rutger J HassinkRHRutger J Hassink
- Thierry X WildberghTWThierry X Wildbergh
- Pieter A DoevendansPDPieter A Doevendans
- Joris JaspersJJJoris Jaspers
- René van EsRERené van Es
Open Access
- 7 July 2023
- journal article
- review article
- Published by JMIR Publications Inc. in JMIR Cardio
- Vol. 7, e44003
Abstract
Background: Electrocardiograms (ECGs) are used by physicians to record, monitor, and diagnose the electrical activity of the heart. Recent technological advances have allowed ECG devices to move out of the clinic and into the home environment. There is a great variety of mobile ECG devices with the capabilities to be used in home environments. Objective: This scoping review aimed to provide a comprehensive overview of the current landscape of mobile ECG devices, including the technology used, intended clinical use, and available clinical evidence. Methods: We conducted a scoping review to identify studies concerning mobile ECG devices in the electronic database PubMed. Secondarily, an internet search was performed to identify other ECG devices available in the market. We summarized the devices’ technical information and usability characteristics based on manufacturer data such as datasheets and user manuals. For each device, we searched for clinical evidence on the capabilities to record heart disorders by performing individual searches in PubMed and ClinicalTrials.gov, as well as the Food and Drug Administration (FDA) 510(k) Premarket Notification and De Novo databases. Results: From the PubMed database and internet search, we identified 58 ECG devices with available manufacturer information. Technical characteristics such as shape, number of electrodes, and signal processing influence the capabilities of the devices to record cardiac disorders. Of the 58 devices, only 26 (45%) had clinical evidence available regarding their ability to detect heart disorders such as rhythm disorders, more specifically atrial fibrillation. Conclusions: ECG devices available in the market are mainly intended to be used for the detection of arrhythmias. No devices are intended to be used for the detection of other cardiac disorders. Technical and design characteristics influence the intended use of the devices and use environments. For mobile ECG devices to be intended to detect other cardiac disorders, challenges regarding signal processing and sensor characteristics should be solved to increase their detection capabilities. Devices recently released include the use of other sensors on ECG devices to increase their detection capabilities.Keywords
All Related Versions
This publication has 83 references indexed in Scilit:
- Screening of Paroxysmal Atrial Fibrillation after Ischemic Stroke: 48-Hour Holter Monitoring versus Prolonged Intermittent ECG RecordingISRN Stroke, 2014
- Feasibility and cost-effectiveness of stroke prevention through community screening for atrial fibrillation using iPhone ECG in pharmaciesThrombosis and Haemostasis, 2014
- Wearable seismocardiography: Towards a beat-by-beat assessment of cardiac mechanics in ambulant subjectsAutonomic Neuroscience, 2013
- iPhone ECG application for community screening to detect silent atrial fibrillation: A novel technology to prevent strokeInternational Journal of Cardiology, 2013
- Use of a Noninvasive Continuous Monitoring Device in the Management of Atrial Fibrillation: A Pilot StudyPacing and Clinical Electrophysiology, 2012
- Improved screening for silent atrial fibrillation after ischaemic strokeEP Europace, 2012
- Short-term ECG for out of hospital detection of silent atrial fibrillation episodesScandinavian Cardiovascular Journal, 2009
- Recommendations for the Standardization and Interpretation of the ElectrocardiogramCirculation, 2007
- ABC of clinical electrocardiography: Introduction. I---Leads, rate, rhythm, and cardiac axisBMJ, 2002
- Recommendations for standardization and specifications in automated electrocardiography: bandwidth and digital signal processing. A report for health professionals by an ad hoc writing group of the Committee on Electrocardiography and Cardiac Electrophysiology of the Council on Clinical Cardiology, American Heart Association.Circulation, 1990