MSCT-fluoroscopy fusion imaging for guidance in percutaneous bicaval valve implantation for transcatheter treatment of tricuspid regurgitation

Authors:

Abstract
A 77-year-old woman, with atrial fibrillation and history of stroke, presented with symptomatic torrential tricuspid regurgitation (TR). Echocardiogram and right heart catheterization confirmed severe TR with slightly elevated pulmonary artery pressure. Surgical risk was considered high (Society of Thoracic Surgery score 7.21%) and was not a good candidate for other percutaneous tricuspid repair techniques. Based on multislice computed tomography (MSCT)-fluoroscopy fusion imaging (HeartNavigator system; Philips Healthcare), two self-expanding percutaneous heart valves were custom-made with 10–20% oversizing to fit the anticipated implantation site in the superior vein cave (SVC) and inferior vein cave (IVC). The IVC valve was designed with the upper segment protruding into the right atrium with...