Abstract
Monitoring bodily pressures provide valuable diagnostic and prognostic information. In particular, long-term measurement through implantable sensors is highly desirable in situations where percutaneous access can be complicated or dangerous (e.g., intracranial pressure in hydrocephalic patients). In spite of decades of progress in the fabrication of miniature solid-state pressure sensors, sensor drift has so far severely limited their application in implantable systems. In this paper, we report on a universal packaging technique for reducing the sensor drift. The described method isolates the pressure sensor from a major source of drift, i.e., contact with the aqueous surrounding environment, by encasing the sensor in a silicone-filled medical-grade polyurethane balloon. In-vitro soak tests for 100 days using commercial micromachined piezoresistive pressure sensors demonstrate a stable operation with the output remaining within 1.8 cmH2O (1.3 mmHg) of a reference pressure transducer. Under similar test conditions, a non-isolated sensor fluctuates between 10 and 20 cmH2O (7.4–14.7 mmHg) of the reference, without ever settling to a stable operation regime. Implantation in Ossabow pigs demonstrate the robustness of the package and its in-vivo efficacy in reducing the baseline drift.Keywords
Funding Information
- ICTSI NIH/NCRR (RR025761)
- NIDDK DiaComp Pilot (13GHSU260)
This publication has 8 references indexed in Scilit:
- Chronically Implanted Pressure Sensors: Challenges and State of the FieldSensors, 2014
- UP-link: An ultra-low power implantable wireless system for long-term ambulatory urodynamics2014
- Intracranial Pressure and Brain Monitoring XIV2012
- Biomedical Sensors and Instruments2011
- Wireless pulmonary artery haemodynamic monitoring in chronic heart failure: a randomised controlled trialThe Lancet, 2011
- Long-term testing of an intracranial pressure monitoring deviceJournal of Neurosurgery, 2000
- Validation of a telemetry system for long-term measurement of blood pressureJournal of Applied Physiology, 1996