Spatial variations in prostate tissue histology as measured by a tactile resonance sensor
Authors:
- Ville JalkanenVJVille Jalkanen
- Britt M AnderssonBABritt M Andersson
- Anders BerghABAnders Bergh
- Börje LjungbergBLBörje Ljungberg
- Olof A LindahlOLOlof A Lindahl
- 28 September 2007
- journal article
- Published by IOP Publishing in Physiological Measurement
- Vol. 28 (10) , 1267-1281
Abstract
In recent years, tactile sensors based on piezoelectric resonance sensor technology have been used for medical diagnosis where the sensor's stiffness-measuring properties can reflect tissue pathology. The change in the frequency of the resonating system and the change in force when contact is made with tissue are used as a stiffness parameter. Earlier stiffness measurements of prostate tissue in vitro demonstrate variations related to tissue composition. In this study, measured stiffness from two human prostate specimens was compared to histological composition of prostate tissue below and around the measurement points. Tissue stiffness was measured with the resonance sensor system. Tissue composition was measured at four different depths in the tissue specimen using a microscopic-image-based morphometrical method. With this method, the proportion of tissue types was determined at the points of intersections in a circular grid on the images representing each measurement point. Numerical values were used for weighting the tissue proportions at different depths in the tissue specimen. For an impression depth of 1.0 mm, the sensing depth in this study was estimated to be 3.5-5.5 mm. Stiffness variations due to horizontal tissue variations were investigated by studying the dependence of the size of the circular grid area relative to the contact area of the sensor tip. The sensing area (grid radius) was estimated to be larger than the contact area (contact radius) between the sensor tip and the tissue. Thus, the sensor tip registers spatial variations in prostate tissue histology, both directly below and lateral to the tip itself. These findings indicate that tumours around the sensor tip could be detected, which in turn supports the idea of a future resonance-sensor-based clinical device for detecting tumours and for guiding biopsies.Keywords
This publication has 13 references indexed in Scilit:
- Resonance sensor measurements of stiffness variations in prostate tissuein vitro—a weighted tissue proportion modelPhysiological Measurement, 2006
- Prostate tissue stiffness as measured with a resonance sensor system: a study on silicone and human prostate tissue in vitroMedical & Biological Engineering & Computing, 2006
- Applanation resonance tonometry for intraocular pressure in humansPhysiological Measurement, 2004
- Real time robotic tactile sensor system for the determination of the physical properties of biomaterialsSensors and Actuators A: Physical, 2004
- Evaluation of applanation resonator sensors for intra-ocular pressure measurement: Results from clinical andin vitro studiesMedical & Biological Engineering & Computing, 2003
- Cancer burden in the year 2000. The global pictureEuropean Journal of Cancer, 2001
- Objective evaluation of liver consistency to estimate hepatic fibrosis and functional reserve for hepatectomy1Journal of the American College of Surgeons, 2000
- The stiffness of lymph nodes containing lung carcinoma metastasesCancer, 1997
- New tactile sensor like the human hand and its applicationsSensors and Actuators A: Physical, 1992
- Contact Mechanics1985