Multiple Probe Frequency Acoustic Reflex Measurements
Authors:
- 1 January 1979
- journal article
- case report
- Published by Taylor & Francis in Scandinavian Audiology
- Vol. 8 (4) , 233-239
Abstract
Acoustic reflex impedance measurements are made on forty-seven normal and two otosclerotic ears using fixed probe frequencies from 220 to 2000 Hz. In general, the reflex increases the middle ear impedance up to 700 Hz probe frequency and decreases it thereafter. It is postulated that this characteristic is related to removal of the cochlear fluid resistance due to stapes decoupling. At one probe frequency in the range 500--900 Hz, the reflex of 50% of the normal ears has either a diphasic pattern or shows no change in the impedance vector length. The accompanying phase change is always monophasic. In the otosclerotic ear the diphasic reflex is present for every probe frequency. These factors should be taken into account when probe frequencies greater than 220 Hz are used for reflex measurement.Keywords
This publication has 8 references indexed in Scilit:
- An Experimental Study on Diphasic Acoustic Reflex Patterns in Normal EarsScandinavian Audiology, 1976
- Acoustic Impedance Measured on Ears with Normal and Diphasic Impedance ChangesScandinavian Audiology, 1975
- On Irregular Acoustic Reflex PatternsScandinavian Audiology, 1975
- Acoustic Middle Ear Muscle Reflexes in Patients With OtosclerosisJAMA Otolaryngology–Head & Neck Surgery, 1973
- Diphasic impedance change and its applicability in clinical workActa Oto-Laryngologica, 1970
- A Quantitative Study of the Effect of the Acoustic Stapedius Reflex on Sound transmission Through the Middle Ear of ManActa Oto-Laryngologica, 1968
- Dynamics of the Acoustic Reflex: Phenomenological AspectsThe Journal of the Acoustical Society of America, 1964
- Network Model of the Middle EarThe Journal of the Acoustical Society of America, 1961