Effect of Silver Annealing Conditions on the Performance of Electrolytic Silver/Silver Chloride Electrodes used in Harned Cell Measurements of pH
Open Access
Abstract
We have studied the long and short term stability of electrolytic Ag/AgCl electrodes fabricated from Ag wire that has been subjected to a range of different annealing conditions. At elevated temperatures, the presence of oxygen during the annealing process has been shown to be detrimental to the performance of electrodes produced. This phenomenon has been attributed to the dissolution of oxygen in the Ag lattice leading to structural changes in the Ag/AgCl electrode material. Electrodes prepared from Ag wire annealed in the absence of oxygen have shown no appreciable change in performance throughout the temperature range employed. This work has resulted in an improved understanding of the optimum annealing conditions required for Ag used in the preparation of electrolytic Ag/AgCl reference electrodes. This work has positive implications for the accuracy of Harned cell measurements of pH.Keywords
This publication has 14 references indexed in Scilit:
- Meeting the Requirements of the Silver/Silver Chloride Reference ElectrodeJournal of Solution Chemistry, 2009
- Long-term equilibrium potential and electrochemical impedance study of Ag/AgCl electrodes used in Harned Cell measurements of pHAccreditation and Quality Assurance, 2009
- Effect of Structural Design of Silver/Silver Chloride Electrodes on Stability and Response Time and the Implications for Improved Accuracy in pH MeasurementSensors, 2009
- The microporous structure of silver/silver chloride electrodes and the implications for Harned cell operationAccreditation and Quality Assurance, 2005
- Observation of a combined dilution and salting effect in buffers under conditions of high dilution and high ionic strengthAccreditation and Quality Assurance, 2003
- Measurement of pH. Definition, standards, and procedures (IUPAC Recommendations 2002)Pure and Applied Chemistry, 2002
- Determination of pH: Theory and PracticeJournal of the Electrochemical Society, 1973
- Preparation and Reproducibility of the Thermal-Electrolytic Silver-Silver Chloride ElectrodeJournal of the Electrochemical Society, 1957
- Potentiometric method for the accurate measurement of hydrogen-ion activityJournal of Research of the National Bureau of Standards, 1939
- THE ELECTROMOTIVE FORCES OF UNI-UNIVALENT HALIDES IN CONCENTRATED AQUEOUS SOLUTIONSJournal of the American Chemical Society, 1929