Abstract
A microwave imaging method that directly incorporates S-parameters into its objective function is presented. S-parameters are extracted from a time-harmonic finite-element model of an imaging system using an efficient thin-wire antenna sub-grid technique. The microwave imaging system considered is an air-based quasi-resonant imaging chamber that is excited with monopole antennas constructed using actual thin wires. Forward modelling results demonstrate the thin-wire model is accurate over the broad band of frequencies tested. A modified contrast source inversion algorithm that incorporates the measured and modelled S-parameters within its objective function is used to reconstruct the complex-valued permittivity of a simple oil-glycerin-water based breast phantom. Image accuracy metrics demonstrate that single frequency experimental inversion results using the thin-wire antenna model and S-parameter objective function improve tumour detection and artifact reduction for the tested breast phantom.Keywords
Funding Information
- Canadian Foundation for Innovation, Research Manitoba
- Research Alliance of Canada
- NSERC Graduate Student Research Award
This publication has 26 references indexed in Scilit:
- Assessing a Microwave Imaging System for Brain Stroke Monitoring via High Fidelity Numerical ModellingIEEE Journal of Electromagnetics, RF and Microwaves in Medicine and Biology, 2021
- Integrating prior information into microwave tomography Part 1: Impact of detail on image qualityMedical Physics, 2017
- Integrating prior information into microwave tomography part 2: Impact of errors in prior information on microwave tomography image qualityMedical Physics, 2017
- Numerical Modeling and High-Speed Parallel Computing: New Perspectives on Tomographic Microwave Imaging for Brain Stroke Detection and MonitoringIEEE Antennas and Propagation Magazine, 2017
- Real-time Microwave Imaging of Differential Temperature for Thermal Therapy MonitoringIEEE Transactions on Biomedical Engineering, 2014
- 2D Fused image reconstruction approach for microwave tomography: a theoretical assessment using the FDTD modelComputer Methods in Biomechanics and Biomedical Engineering: Imaging & Visualization, 2013
- Analysis of Incident Field Modeling and Incident/Scattered Field Calibration Techniques in Microwave TomographyIEEE Antennas and Wireless Propagation Letters, 2011
- Finite-element contrast source inversion method for microwave imagingInverse Problems, 2010
- 3D Image Reconstruction in Microwave Tomography Using an Efficient FDTD ModelIEEE Antennas and Wireless Propagation Letters, 2009
- An unconditionally stable subcell model for arbitrarily oriented thin wires in the FETD methodIEEE Transactions on Antennas and Propagation, 2003