Correction of Brillouin linewidths measured by multipass Fabry-Perot spectroscopy
Authors:
- 1 May 1977
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 16 (5) , 1404-1407
Abstract
The sharper instrumental profile of the multipassed Fabry-Perot etalon introduces distortions to Brillouin spectra which may differ considerably from those introduced by the single-pass etalon. The half-width at half-height of the observed profile is calculated for five-pass and three-pass spectrometers. Phonon spectral widths may be extracted from the graphs and appropriate relationships are presented. These calculations are valid for well aligned etalons when the phonon lineshape is Lorentzian and cover a wide range of instrumental and phonon spectral widths.Keywords
This publication has 7 references indexed in Scilit:
- Multi-pass Fabry-Perot spectroscopy of polymersPolymer, 1976
- Brillouin scattering and the glass‐rubber relaxation in amorphous polymersJournal of Polymer Science Part C: Polymer Letters, 1975
- Brillouin spectra of CaF2 microcrystals using a stable 3-pass Fabry-Perot interferometerApplied Physics Letters, 1973
- Brillouin scattering study of SbSI using a double-passed, stabilised scanning interferometerOptics Communications, 1970
- Brillouin Scattering and Relaxation in LiquidsThe Journal of the Acoustical Society of America, 1968
- Instrumental Effects on Brillouin Line ShapesThe Journal of the Acoustical Society of America, 1968
- An algorithm for the machine calculation of complex Fourier seriesMathematics of Computation, 1965