Abstract
We studied the effect of an adhesive external nasal dilator strip (ENDS) on external nasal geometry in 20 healthy Caucasian adults (10 men, 10 women; age 21–45 yr). The recoil force exerted by ENDS was estimated by bending the device (n = 10) with known weights. In the horizontal direction, a small/medium-sized ENDS in situ exerted a unilateral recoil force of 21.4–22.6 g. Application of ENDS resulted in a displacement of the lateral nasal vestibule walls that had both anterosuperior and horizontal components and that was maintained over an 8-h period. The resultant unilateral nasal vestibule wall displacement at the tip of the device was at 47.6 ± 2.0° to the horizontal (as related to the plane of the device when in situ) and had a magnitude of 3.5 ± 0.1 mm. ENDS increased external nasal cross-sectional area by 23.0–65.3 mm2. Nasal vestibule wall compliance was estimated at 0.05–0.16 mm/g. Thus ENDS applies a relatively constant abducting force irrespective of nasal width. Variable responsiveness to ENDS may be related to differences in elastic properties of the nasal vestibule wall.

This publication has 10 references indexed in Scilit: