Efficacy and Safety of PolarX for Pulmonary Vein Isolation in Atrial Fibrillation: A Systematic Review and Meta‐Analysis

Authors:

Abstract
Background: The efficacy and safety of PolarX cryoballoon for pulmonary vein isolation (PVI) among patients with atrial fibrillation (AF) is unclear. Methods: PubMed, Embase, Cochrane Central, and ClinicalTrials.gov were systematically searched for studies comparing the PolarX vs. standard cryoballoon (CB‐2/CB‐4). Odds ratios (ORs) and mean differences (MDs) with 95% confidence intervals (95% CIs) were computed for binary and continuous outcomes, respectively. Results: This systematic review and meta‐analysis included 2 randomized controlled trials and 18 observational studies with a total of 7107 patients, of whom 3.188 (44.8%) underwent ablation with PolarX. When compared to standard cryoablation, PolarX had similar incidence of atrial arrhythmias recurrence (OR 1.04; 95% CI 0.82–1.32; p = 0.73), acute PVI success (OR 0.91; 95% CI 0.31–2.68; p = 0.98), and procedural duration (MD 3.41; 95% CI ‐2.89 to 9.71; p = 0.29). However, PolarX was associated with higher persistent phrenic nerve palsy (OR 2.17; 95% CI 1.26–3.74; p = 0.005) and a lower nadir temperature in all veins. Additionally, in a subgroup analysis of randomized trials and propensity score matching studies, similar findings were noted for recurrence of atrial arrhythmias. Conclusion: In patients with AF undergoing PVI, PolarX had similar acute PVI success, procedural duration, and recurrence of atrial arrhythmias as compared with standard cryoballoon technology, however, was associated with a lower nadir temperature, and higher rates of persistent nerve palsy.

This publication has 47 references indexed in Scilit: