Complications of Implantable Cardioverter Defibrillator Therapy in 440 Consecutive Patients
Authors:
- 15 September 2005
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 28 (9) , 926-932
Abstract
Background: Although more than 150,000 implantable cardioverter defibrillators (ICDs) are implanted yearly worldwide, only few studies systematically examined complications of ICD therapy in large patient cohorts. Methods: We prospectively analyzed ICD‐related complications in 440 consecutive patients who underwent first implantation of an ICD system for primary or secondary prevention of sudden cardiac death within the last 10 years at our institution. All study patients received pectoral nonthoracotomy ICD lead systems with the exception of one patient who had an artificial tricuspid valve. Results: During 46 ± 37 months follow‐up, 136 of 440 patients (31%) experienced at least one complication including implant procedure‐related complications in 43 patients (10%), ICD generator‐related complications in 28 patients (6%), lead‐related complications in 52 patients (12%), and inappropriate shocks in 54 patients (12%). The most serious complications included one perioperative death due to heart failure (0.2%), two ICD system infections necessitating device removal (0.5%) and two perioperative cerebrovascular strokes (0.5%). Conclusions: We conclude that more than one quarter of ICD patients experience complications during a mean follow‐up of almost 4 years, although serious complications such as intraoperative death or ICD system infections are rare in patients with nonthoracotomy ICD systems. Recognition of these complications is the prerequisite for advances in ICD technology and management strategies to avoid their recurrence.Keywords
This publication has 31 references indexed in Scilit:
- Death Due to an Implantable Cardioverter DefibrillatorJournal of Cardiovascular Electrophysiology, 2004
- Mortality, Morbidity, and Complications in 3,344 Patients with Implantable Cardioverter Defibrillators:Pacing and Clinical Electrophysiology, 2003
- Prophylactic Implantation of a Defibrillator in Patients with Myocardial Infarction and Reduced Ejection FractionNew England Journal of Medicine, 2002
- The Cardiac Insufficiency Bisoprolol Study II (CIBIS-II): a randomised trialThe Lancet, 1999
- Complications Associated with Pectoral Implantation of Cardioverter DefibrillatorsPacing and Clinical Electrophysiology, 1997
- Complications associated with pectoral cardioverter-defibrillator implantation: Comparison of subcutaneous and submuscular approachesJACC, 1996
- Multicenter experience with a pectoral unipolar implantable cardioverter-defibrillatorJACC, 1996
- Multicenter Experience With a Pectoral Unipolar Implantable Cardioverter-DefibrillatorJACC, 1996
- Postoperative lead-related complications in patients with nonthoracotomy defibrillation lead systemsJACC, 1995
- Low Incidence of Lead Related Complications Associated with Nonthoracotomy Implantable Cardioverter Defibrillator SystemsPacing and Clinical Electrophysiology, 1995