Present state and issues in IORT Physics
Open Access
- 27 January 2017
- journal article
- review article
- Published by Springer Nature in Radiation Oncology
- Vol. 12 (1) , 1-30
Abstract
Literature was reviewed to assess the physical aspects governing the present and emerging technologies used in intraoperative radiation therapy (IORT). Three major technologies were identified: treatment with electrons, treatment with external generators of kV X-rays and electronic brachytherapy. Although also used in IORT, literature on brachytherapy with radioactive sources is not systematically reviewed since an extensive own body of specialized literature and reviews exists in this field. A comparison with radioactive sources is made in the use of balloon catheters for partial breast irradiation where these are applied in almost an identical applicator technique as used with kV X-ray sources. The physical constraints of adaption of the dose distribution to the extended target in breast IORT are compared. Concerning further physical issues, the literature on radiation protection, commissioning, calibration, quality assurance (QA) and in-vivo dosimetry of the three technologies was reviewed. Several issues were found in the calibration and the use of dosimetry detectors and phantoms for low energy X-rays which require further investigation. The uncertainties in the different steps of dose determination were estimated, leading to an estimated total uncertainty of around 10-15% for IORT procedures. The dose inhomogeneity caused by the prescription of electrons at 90% and by the steep dose gradient of kV X-rays causes additional deviations from prescription dose which must be considered in the assessment of dose response in IORT.Keywords
This publication has 116 references indexed in Scilit:
- Can we improve the dose distribution for single or multi-lumen breast balloons used for Accelerated Partial Breast Irradiation?Journal of Contemporary Brachytherapy, 2013
- In vivo dosimetry with radiochromic films in low‐voltage intraoperative radiotherapy of the breastMedical Physics, 2012
- Radiation protection for an intra-operative X-ray deviceBritish Journal of Radiology, 2011
- Cone-beam CT with a flat-panel detector: From image science to image-guided surgeryNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2011
- Monte Carlo simulation of electron beams generated by a 12 MeV dedicated mobile IORT acceleratorPhysics in Medicine & Biology, 2011
- Accelerated Partial Breast Irradiation (APBI): A review of available techniquesRadiation Oncology, 2010
- Kypho-IORT - a novel approach of intraoperative radiotherapy during kyphoplasty for vertebral metastasesRadiation Oncology, 2010
- Is Electron Beam Intraoperative Radiotherapy (ELIOT) Safe in Pregnant Women with Early Breast Cancer? In Vivo Dosimetry to Assess Fetal DoseAnnals of Surgical Oncology, 2009
- Investigation of the usability of conebeam CT data sets for dose calculationRadiation Oncology, 2008
- Surgical technique of intraoperative radiation therapy with electrons (ELIOT) in breast cancer: A lesson learned by over 1000 proceduresSurgery, 2006