Validity of a Tympanic Thermometer and Thermal Imaging Camera for Measuring Core and Skin Temperature during Exercise in the Heat
- 17 September 2019
- journal article
- research article
- Published by Taylor & Francis in Measurement in Physical Education and Exercise Science
- Vol. 24 (1) , 49-55
Abstract
This study compared criterion to minimally invasive, practical measures of core (Tc) and skin (Tsk) temperature during 30 min of moderate-intensity cycle exercise in a heat chamber (35°C, 60% rH). Tc was monitored using a core temperature pill (Tc(Pill)) and tympanic thermometer (Tc(Tymp)) during rest, exercise, and recovery in 15 participants. Tsk was monitored using hard-wired skin thermistors attached to a data logger (Tsk(T)) and a thermal imaging camera (Tsk(IR)) in 11 participants. Tc measurement resulted in no significant difference (p > 0.05), a mean bias of 0.1°C, coefficient of variation (CV%) of 1.0%, and correlation of r = 0.74 between devices. Tsk measurement resulted in a significant difference (p = 0.01), a mean bias of 0.6°C, CV% of 2.3%, and correlation of r = 0.61 between devices. Tc(Tymp) demonstrated acceptable agreement with Tc(Pill); however, caution is advised when using Tsk(IR) to give accurate measures of Tsk during exercise.Keywords
This publication has 27 references indexed in Scilit:
- Accuracy of tympanic temperature measurement using an infrared tympanic membrane thermometerBMC Research Notes, 2013
- Telemetry pill versus rectal and esophageal temperature during extreme rates of exercise-induced core temperature changePhysiological Measurement, 2012
- The use of thermal imaging to evaluate body temperature changes of athletes during training and a study on the impact of physiological and morphological factors on skin temperatureHuman Movement, 2012
- The effect of skin thermistor fixation method on weighted mean skin temperaturePhysiological Measurement, 2011
- Validity and Reliability of Devices That Assess Body Temperature During Indoor Exercise in the HeatJournal of Athletic Training, 2009
- Comparison of infrared versus contact thermometry for measuring skin temperature during exercise in the heatPhysiological Measurement, 2007
- Rectal, telemetry pill and tympanic membrane thermometry during exercise heat stressJournal of Thermal Biology, 2007
- The ingestible telemetric body core temperature sensor: a review of validity and exercise applicationsBritish Journal of Sports Medicine, 2006
- Assessment of an infra-red non-contact sensor for routine skin temperature monitoring: a preliminary studyJournal of Medical Engineering & Technology, 1992